Application
|
25
For IEC/EN 31000-4-3 type testing, the antenna tip can be placed at any
distance between one and three meters from the EUT as long as the front face
plane is illuminated according to the -0, +6 dB uniform field specifications.
It is usually necessary to place RF absorbing material between the EUT and
antenna to suppress ground plane reflection to ensure the field uniformly, or to
conduct the immunity test in a fully-lined anechoic room. In general, closer
distances require less power to create a given field strength.
Содержание 3148B
Страница 1: ...Model 3148B Log Periodic Dipole Array Antenna User Manual ...
Страница 6: ...vi This page intentionally left blank ...
Страница 10: ...10 Introduction This page intentionally left blank ...
Страница 14: ...14 Specifications This page intentionally left blank ...
Страница 17: ...Mounting Instructions 17 Shown mounted onto a 4 TR ...
Страница 22: ...22 Mounting Instructions This page intentionally left blank ...
Страница 26: ...26 Application This page intentionally left blank ...
Страница 27: ...Typical Data 27 6 0 Typical Data Antenna Factor ...
Страница 28: ...28 Typical Data Gain ...
Страница 29: ...Typical Data 29 VSWR ...
Страница 30: ...30 Typical Data Half Power Beamwidth ...
Страница 31: ...Typical Radiation Patterns 31 7 0 Typical Radiation Patterns 400 MHz ...
Страница 32: ...32 Typical Radiation Patterns 500 MHz 600 MHz ...
Страница 33: ...Typical Radiation Patterns 33 700 MHz 800 MHz ...
Страница 34: ...34 Typical Radiation Patterns 900 MHz 1000 MHz ...
Страница 35: ...Typical Radiation Patterns 35 1100 MHz 1200 MHz ...
Страница 36: ...36 Typical Radiation Patterns 1300 MHz 1400 MHz ...
Страница 37: ...Typical Radiation Patterns 37 1500 MHz 1600 MHz ...
Страница 38: ...38 Typical Radiation Patterns 1700 MHz 1800 MHz ...
Страница 39: ...Typical Radiation Patterns 39 1900 MHz 2000 MHz ...
Страница 40: ...40 Typical Radiation Patterns This page intentionally left blank ...