Model 2090 Multi-Device Controller
Parameter List
P1
Device Type
: (0 = rotational positioner, 1 = linear positioner)
P2
Device Sub-Type: (0 = Standard Turntable/Tower, 1 = Air Flotation
Turntable/Bore Sight Tower, 2 = Two Speed Turntable)
P3
SCAN cycle count
: (1-999 complete scans, 0 = infinite scan)
P4
Polarization offset
: (
±
50 cm) -or-
Bore Sight separation distance
: (3, 10, or 30 m)
P5
Bore sight correction table
: (0 = standard, 1 = alternate) -or-
Turntable rotation mode
: (0 = Continuous, 1 = Non-continuous)
P8
Motor reverse delay
: (0.1 – 99.9 seconds)
P9
GPIB Address
: (0-30)
B1
Bit 0, STEP limits
– (0 = INC/DEC limited by hard limits, 1 = INC/DEC limited by soft limits)
Bit 1, Quick Stop
– (0 = normal deceleration, 1 = fast deceleration (variable speed only))
C
Encoder calibration
: Encoder counts per meter/revolution (1-9999). Typical values are:
Towers: Default = 2000, 2075 = 1620; Turntables: Default = 3600, 2065 = 3665, 2081 = 3608,
2088 = 4500 (4800*), 5901 = 6000 (6400*), MAPS = 4500. * Early revisions.
S0
STEP speed setting
: (-1 = disabled, 0-255 (variable speed devices only)
S1-S8
Speed setting
: (0-255 (variable speed devices only)
PCLr
Clear Parameters
: Restores this device to factory default.
OSCLr
Clear Overshoot
: Clears overshoot compensation values for this device.
OC
Overshoot Compensation Enable/Disable
: (
On
= enabled,
OFF
= disabled)
ZERO
Zero Reference
: Initiates a zero pulse reference scan on equipped devices.
BASE
Motor base firmware revision
: For informational purposes only.
ACC
Acceleration
: Time in seconds for motorbase to reach full speed (variable speed devices only)
TE P
Temperature monitor
: Displays motor and control board temperature.
SND
Motorbase sound setting
: (0 = No sounds, 1 = remote control only, 2 = errors enabled, 3 = all
sounds enabled).
© ETS-Lindgren, April 2006
Revision G– P#399199
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com