FEATURES
START
LCD
POWER
(ON/MEM)
Lens
Battery
Cover
This non-contact digital forehead thermometer is intended for the intermittent measurement of
human body temperature in people of all ages.
OPERATING INSTRUCTIONS FOR FOREHEAD USE
-
Press to turn on the instrument. Hold the
instrument within 4 cm from the centre of the forehead and press and hold ‘Start’ until you hear one
beep (unless muted) to indicate the reading is complete. Readings should take approximately one
second. The instrument will automatically turn off after 60 seconds. Manually turn the instrument
off by pressing and holding the for five seconds.
The temperature displayed is the “oral equivalent” value. If the temperature is above 37.5 °C
(99.5 °F) you will hear three short beeps followed by one long beep.
FOR BEST RESULTS
-
Keep the forehead area clean. Keep away from sweat, cosmetics, and scars
while taking measurements. Please remain in a stable environment for five minutes, and avoid
exercising, bathing, or showering for 30 minutes, which can artificially raise your temperature.
Non-contact
FOREHEAD THERMOMETER
Operating Instructions
Product code: 801-590
STORAGE & CLEANING-
The instrument should be stored at room temperature, away from
liquids and direct sunlight. If there are any temperature differences between the place where
the instrument is stored and where you are going to measure, please allow the instrument to
stabilize to the ambient temperature for at least 15 minutes. Holding the instrument too long
may cause a higher ambient temperature reading. This could make the body temperature
measurement lower than usual.
The sensor lens is the most delicate part of the instrument and should be kept clean at all times.
Care should be taken when cleaning the lens to avoid damage. Use only a soft cloth or cotton
swab with medical alcohol, allowing the lens to dry fully before using the instrument again.
SPECIFICATIONS
GUARANTEE
- This instrument carries a one-year guarantee against defects in either components
or workmanship. During this period, products that prove to be defective will, at the discretion of
ETI, be either repaired or replaced without charge. The product guarantee does not cover damage
caused by fair wear and tear, abnormal storage conditions, incorrect use, accidental misuse, abuse,
neglect, misapplication or modification. Full details of liability are available within ETI’s Terms
& Conditions of Sale at etiltd.com/terms. In line with our policy of continuous development, we
reserve the right to amend our product specification without prior notice.
Forehead range
Surface range
Operating range
Storage range
Transport. temp
Atmospheric pressure
Forehead accuracy
Surface accuracy
Resolution
Units
Response time
Auto-Off
Battery
Dimensions
Weight
Expected service life
34 to 42.2 °C (93.2 to 108 °F)
- 22 to 80 °C (- 7.6 to 176 °F)
10 to 40 °C (50 to 104 °F ) 15 to 85 %rh
- 20 to 50 °C (- 4 to 122 °F) , < 85 %rh
Shal
l be less than 70 °C (158 °F), 95 %rh
800 to 1013 hPa
±0.2 °C (0.4 °F ) within 35 to 42 °C
(95 to 107.6 °F) Otherwise ±0.3 °C (0.5 °F)
±0
.3 °C (0.5 °F) with 22 to 42.2 °C (17.6 to 108 °F)
Otherwise ±4% or ±2 °C (±4 °F),
whichever is greater
0.1 °C / °F
°C / °F
1 second
60 seconds
2 x 1.5 volt AAA - 3000 hours
40.2 x 48 x 158 mm
100 grams
4 years