Gripping Surface

Anvil Half

Locking Rib

Anvil Fork

Staple Retaining Cap

Alignment/Locking Lever

Alignment Slot

Cartridge/Reload Fork

Staple Height Selector

Cartridge/Reload Half 

Firing Knob

Gripping Surface

Alignment Tab

Gripping Surface

2

4

1

3

5

Steps to Use

Models available: (55mm, 75mm)

NTLC55, NTLC75

Reloads available:

SR55, SR75

Verify compatibility of all instruments and accessories prior 
to using the instrument. For complete product details, see 
Instructions for Use.

Using sterile technique, remove the instrument from the 
package. To avoid damage, do not flip the Cartridge/Reload 
into the sterile field.

If instrument is not separated into halves, separate instrument by 
completely disengaging the Alignment/Locking Lever.

Load the instrument by inserting the selectable Cartridge/Reload 
by placing the alignment tabs into the alignment slots and pivoting 
the selectable Cartridge/Reload onto the Cartridge/Reload Fork. 
Snap the Cartridge/Reload into position.

Prior to placing instrument across the tissue, select the desired closed staple height. In order to select the desired staple height, push 
the Staple Height Selector so that it is aligned with desired color indicator, Blue 1.5 mm, Gold 1.8 mm or Green 2.0 mm. If tissue cannot be 
comfortably compressed to the closed staple height, or easily compresses to less than the closed staple height, either the setting or the 
tissue is contraindicated for the selected staple size or setting.

CAUTION:

 Careful consideration should be given to any pre-surgical treatment the patient may have undergone, which may require 

alterations to surgical technique or alternative surgical procedures.

CAUTION:

 Tissue thickness should be carefully evaluated before firing any stapler.

Remove Staple Retaining Cap by grasping the edge of the Staple 
Retaining Cap and lift straight up from the Cartridge/Reload. 
Discard the Staple Retaining Cap. 

Note:

 The Staple Retaining Cap ensures proper staple orientation 

and protects the staple leg points during shipping and 
transportation.