ETHICON Linear Cutter
*
Steps to Use
*Ethicon Endo-Surgery™ Linear Cutter
Страница 1: ...ETHICON Linear Cutter Steps to Use Ethicon Endo Surgery Linear Cutter...
Страница 2: ...ridge Reload Fork Snap the Cartridge Reload into position Prior to placing instrument across the tissue select the desired closed staple height In order to select the desired staple height push the St...
Страница 3: ...tional The Alignment Locking Lever may be moved from the completely opened to the completely closed position CAUTION Ensure that the tissue lies flat between the forks Any bunching of tissue along the...
Страница 4: ...CAUTION The firing stroke must be completed Do not partially fire the instrument Incomplete firing can result in malformed staples incomplete cut line bleeding and leakage from the staple line and or...
Страница 5: ...on Remove the Staple Retaining Cap and discard CAUTION After reloading observe the surface of the new Cartridge If the colored staple drivers are visible replace with another Reload The staple drivers...