Firing the device
1.
Ensure that the tissue lies flat and is positioned properly between the jaws and that no
obstructions (e.g., clips, stents) are within the jaws. Any “bunching” of tissue along the
reload may result in an incomplete staple line.
3.
Pull back the red firing trigger lock.
4.
Pull the exposed trigger to fire (motor will be audible).
Note:
Alternatively, for enhanced tissue compression when using the ECHELON FLEX™
Powered Plus Stapler, the user may pulse the firing trigger throughout the firing cycle until the
transection is completed. Similarly, if the firing sequence is interrupted, the user may pulse the
firing trigger to continue the transection. It is recommended that the pulsing technique not be
initiated until after the knife blade indicator has moved into the closed bottom channel and is
no longer visible to minimize potential of inadvertent lockout.
5.
When the motor stops, release the firing trigger to automatically return the knife to the
home position.
2.
Fully close the closing trigger until it clicks (audible click will occur).
Note:
Holding the jaws in place for 15 seconds after closing and prior to firing may result in
better compression and staple formation.