MANUAL SGM800
81
2.1.15
Menu -tCL-
tCL Theoretic calibration
In this menu, the
Theoretic calibration settings
can be set. Here you can set a
calibration without using a reference weight. For this you only need the specification
sheets of the used loadcells. Options are:
tCL 1
Maximum load loadcells
tCL 2
Sensitivity loadcell 1
tCL 3
Sensitivity loadcell 2
tCL 4
Sensitivity loadcell 3
tCL 5
Sensitivity loadcell 4
Press key 2 <2 sec to enter the theoretic calibration settings.
The following screen is visible:
In
tcL 1
you can set the
maximum loadcell load
. When more then one loadcell is used all
loadcells should have the same maximum load. Press key 2 <2 sec. to set the maximum
load.
The following screen is visible: