ENGLISH
17
Sel ect M E A S U R E M E N T T Y P E P H or R X
The instrument of the eSelect M2 series makes it possible to select the type of measurement to be used, i.e. PH or RX (Redox) in the
same unit. The measurement parameter may be changed from PH to RX and the display and resolutions of the measurement change
accordingly.
◄
►◄
►
By selecting one of the two types of measurement,
programming of the main Menu changes accordingly.
IMPORTANT:
after selecting
M easurem ent Type
PH or RX,
ENSURE
you connect the appropriate electrode.
In case the operator should want
M easurem ent Type
also during operations, proceed as follows:
BASIC
menu > S E T T I N G S
>
M E N U E X P E R T > M E N U S E L E C T I O N
> S E T P O I N T 1
>
Me a s u r e m e n t
T y p e
a
from
EXPERT
menu go into >
M E N U S E L E C T I O N
>
S E T P O I N T 1
>
M e a s u r e m e n t T y p e
a
Press E S C to go back to
M E N U S E L E C T I O N
or press E S C
E S C to go back to
M E A S U R E M E N T D I S P L A Y A
Sel ect th e C H L O R I N E M easurem ent R A N G E
The instrument eSelect M2 PH(RX)-CL offers the option of selecting 4 chlorine measurement ranges and related sensors:
-
Ion-selective membrane sensors with measurement ranges 0-2 Cl ppm (preset); 0-20 Cl ppm; 0-20 Cl ppm (Total Chlorine);
-
Open type amperometric cell (model
CLC
) range 0-10 Cl ppm
After making the selection, the measurement parameters displayed will change accordingly.
IMPORTANT:
after selecting the
CHLORINE SCALE
,
ENSURE
the
adequate
chlorine sensor is connected.
In case the user should wish to change the range via
CHLORINE SCALE
, even during operations:
BASIC
menu > S E T T I N G S >
E X P E R T M E N U > M E N U S E L E C T I O N > S E T P O I N T 1
>
C h l o r i n e S c a l e
o
from
EXPERT
menu go into >
M E N U S E L E C T I O N
>
S E T P O I N T 1
>
S
C h l o r i n e
S c a l e
o
Press E S C to go back to M E N U S E L E C T I O N
or press E S C
E S C
to go back to
M E A S U R E M E N T D I S P L A Y
A
T E M P E R A T U R E setti ngs
Manual temperature offset 0-100°C.
Press
◄►
to change the ambient temperature.
Select AUTOMATIC offsetting for measurements and temperature
offsetting with PT100 sensor.
D A T E A N D TI M E – TI ME R i n R E A L TI M E
◄
►◄
►
Program the date and time by selecting the backlit steps.
Measurement Type
pH
Measurement Type
Rx
SETTINGS
Chlorine Scale
Cl 0-2 p p m
�
MENU BASE
�
SETTINGS
Chlorine Scale
Cl 0-20 ppm
�
MENU BASE
�
SETTINGS
Chlorine Scale
Cl Tot
�
MENU BASE
�
SETTINGS
Chlorine Scale
CLC (0-10 ppm)
�
MENU BASE
�
SETTINGS
Temp. Sens or
Automat ic
SETTINGS
Temperature
25°C
SETTINGS
Temp. Sensor
Manua l
Date
2014 May 14
Time
12:02