Calibration mV - AG-Select B1

From main screen select OK and press PLUS button

X 3 or until “CAL” is shown

Press OK button to go to 1

st

calibration point 650mV

Place clean mV electrode into the buffer solution,mix slowly and 

then wait until display stabilises after around 1 minute. Use the 

PLUS and MINUS buttons to trim the reading to mV 650

Press OK button to confirm and to go to next calibration point for 

“ZERO” mV. To simulate 0 mV use either a shorting plug or using 

a small piece of wire connect the outer body of the BNC 

connector to the inner core section.

Proceed as for CAL1 and press OK button when done and press 

ESC button to navigate back to start up screen 