27
Molift QuickRaiser 205 / www.etac.com
Replace PCB / replace fuse / replace coin battery
Push in emergency stop and discon-
nect battery before any repair or
service.
Remove top cover, see page 26.
Disconnect all cables
Replace Fuse
Change fuse art.2920161
Replace PCB
Remove 2x torx screws (T10), Change PCB art.
2920112
Replace Coin battery
Change battery art
.
2920105. Use a flathead
screwdriver and tilt the battery off. When battery
has been disconnected, the clock must be set in
Molift Service Tool.
On the left hand side meny, there is a button “Set
clock…” This feature sets the clock on the PCB
according to the PC connected. Press “OK” to exit
the feature.
Check after repair
Make sure all functions on control panel
are working (up/down)
Make sure service light is not flashing
yellow
Check settings in Molift Service Tool
Perform and fill out Safety checklist
after service and repair
Parts
Article no.
2920161
Fuse, 10A
2920112
PCB (includes Fuse and battery)
2920103
Battery CR1632 QR205
Содержание Molift QuickRaiser 205
Страница 1: ...TM13001 Rev B 2018 01 03 Molift QuickRaiser 205 EN Technical manual...
Страница 12: ...11 Molift QuickRaiser 205 www etac com Driveline Article no 2910030 6 6 1 4 5 3 2...
Страница 50: ...49 Molift QuickRaiser 205 www etac com Manufacturer s Declaration...
Страница 51: ...50 Molift QuickRaiser 205 www etac com Manufacturer s Declaration...
Страница 52: ...51 Molift QuickRaiser 205 www etac com Manufacturer s Declaration...
Страница 53: ...52 Molift QuickRaiser 205 www etac com Notes...
Страница 54: ...Etac AS Etac Supply Gj vik Hadelandsveien 2 2816 Gj vik Norway Tel 47 4000 1004 molift etac com www etac com...