Functional examination

Test FUNCTION and inspect for wear 
and damage. 
All checkpoints must be checked to 
approve hoist for further use

OK

Not OK

Wheels / castors

Rear wheel / castor brakes

With legs in inner position (par-
allel): Check that the distance 
between legs can not decrease 
when pushed together by hand
For Mover 180 only: Lifting 

column is completely fixed with 

hexagon locking screws tightly 
secured
For Mover 205 and 300 only: 
Lifting column is completely 

fixed when locking handle is 

engaged
Emergency stop button is fully 
operational and interrupts 
power when engaged.  
Reset switch. 

Electrical emergency lowering

Manual emergency lowering

 Hand control (buttons, strain 
relief on hand control cable, 
hook, and battery indicator)
Sling bar coupling (swivel joint 
moves freely and without noise)

Sling bar quick release

If all points so far are “

OK

”, 

hoist shall be tested with SWL.
Perform one lifting cycle with 
weight equal to SWL (see prod-
uct label for value kg/lbs). 
Actuator must not slip when 
loaded (causing lifting arm to 
lower).
Leg spreading mechanism, 
outwards and  inwards with SWL
Perform new visual control; 
Damage, play and deformations 
as described above. Any dam-
aged parts must be repaired or 
replaced and test is repeated 
once more until the hoist 
performs correctly

OK

Not OK

Charger is operational

With fully charged battery 
inserted, after initial cycle, 
battery light is off, and wrench /
service light is green

Accessories:

......................................................................................

......................................................................................

......................................................................................

Performed by

Full name:

Date/Place:

Signature:

 

 Approved without faults

 
 

Next inspection (YYYY / MM): .........../............

 

 Hoist is marked with “Out of order” and sent 
for repair

 

 Hoist is not eligible for repair and taken out 
of service

If periodic inspection reveals any defect, wear 
or other damage that jeopardises the safety of 
the patient the hoist may not be used until the 

deficiency has been eliminated

 

 The owner is notified

Содержание Molift Mover 180

Страница 1: ...ABCDEFGHIJ Product label Mover 180 ID label Mover 180 Serial no Article no Production date SWL value Lifting arm Push handles Motor actuator Emergency stop Hexagon socket screw Legs Electric emergency lowering Chassis Sling bar Battery Lifting column Manual emergency lowering Hand control Wheel castor Rear with brake Hexagon socket screw PI06101 Rev D 2022 06 15 Periodic inspection checklist EN In...

Страница 2: ...XX XXXXXX YYYY MM DD Etac A S Parallelvej 3 DK 8751 Gedved www etac com ABCDEFGHIJ Product label Mover 205 ID label Mover 205 Serial no Article no Production date SWL value Lifting arm Push handles Motor actuator Emergency stop Legs Electric emergency lowering Chassis Sling bar Battery Lifting column Manual emergency lowering Hand control Wheel castor rear with brake Locking handle ...

Страница 3: ...XX XXXXXX YYYY MM DD Etac A S Parallelvej 3 DK 8751 Gedved www etac com ABCDEFGHIJ Product label Mover 300 ID label Mover 300 Serial no Article no Production date SWL value Lifting arm Push handles Motor actuator Emergency stop Locking handle Legs Electric emergency lowering Chassis Sling bar Battery Lifting column Manual emergency lowering Hand control Wheel castor Rear with brake ...

Страница 4: ...nd name of inspector along with additional information Molift recommends adding all observations remarks concerning the hoist in the Molift Service tool for complete history log of the hoist Service light indicator Note colour of service light indicator Status Service light Mode No light Power saving Stand by Green Ready for use Yellow Order service Red Perform service Red and sound when activated...

Страница 5: ...ut noise Sling bar quick release If all points so far are OK hoist shall be tested with SWL Perform one lifting cycle with weight equal to SWL see prod uct label for value kg lbs Actuator must not slip when loaded causing lifting arm to lower Leg spreading mechanism outwards and inwards with SWL Perform new visual control Damage play and deformations as described above Any dam aged parts must be r...

Страница 6: ... the owner s service log book If approved without faults apply inspection label and mark with month and year for next inspection When ordering label use item no 1100306 2 0 2 7 2 0 2 6 2 0 2 5 2 0 2 4 2023 Next periodic inspection Inspection performed by Notes and observations Mark label with month and year for next inspection ...

Отзывы: