Universal interface 

ET System electronic GmbH 

 

 

11 

Instruction Set 

The  IEEE-488.2  standard  demands  several  basic  commands.  Some  commands  may  occur  twice  for  compatibility 
reasons (once in the ETS version and once in the (old) IEEE-488.2 version). The following syntax is used to describe the 
commands: 

[] 

Square brackets 

 optional parameter 

<> 

Angle bracket 

 numerical value 

{} 

Curly bracket 

 selection list 

Vertical line 

 separator within selection list 

 

Example 
 

GTR[,{0|1|2}] means that the command GTR can be used with or without parameters. If a parameter exists, it has to be 1, 2 or 3. Valid 
commands are: GTR 

GTR,1 

GTR,2 

GTR,3 

IA[,<imax>] means that the command IA can be used either with or without parameters. If a parameter exists, it has to be a numerical 
value. 
 
 
 
 

Quick view of commands 

Command 

Description 

Result 

CLS* or CLS 

Clear Status 

Deletes the status byte 

DCL 

Device Clear 

Initialization data reset 

GTL 

Go To Local 

Activates front panel operation 

GTR[,{0|1|2}] 

Go To Remote 

Activates digital interface operation 

IA[,<imax>] 

Set I

max

 

Adjusts current limiting 

ID or *IDN? 

Identification 

Displays identification string 

LLO 

Local Lockout 

Deactivates LOCAL button 

LIMI 

Limit I

Reads maximum adjustable current limitation 

LIMU 

Limit U

Reads maximum adjustable voltage limitation 

MI[,<Nr>] 

Measure I

a

 

Measures present output current 

MU[,<Nr>] 

Measure U

a

 

Measures present output voltage 

 

 

 

*OPT? 

Optional Identification Query 

Displays units current hardware/software version 

OVP[,<U

ovp

>] 

Overvoltage Protection 

Adjusts over voltage protection 

PCx[,<baud>,<parity>, 
<data bits>,<stop bits>, 
<handshake>,<echo>, 
<timeout>] 

Program  

Communication 

Adjusts the interfaces 

RI or *RST 

Reset Instrument 

Resets hardware (no return value) 

SB[,{S|R|1|0}] 

Standby 

Enables/blocks the output 

SS or *PDU 

Save Setup 

Saves previously made channel and interface parameter adjustments (no return 
value) 

STATUS 

Status 

Query of the units’ status (return values in binary format) (also see following 
table) 

STB or *STB? 

Interface Status 

Displays the interface status. 

UA[,<ua>] 

Set U

a

 

Adjusts output voltage (if there are no parameters, present set point is 
displayed) 

Detailed description of commands 

CLS* or CLS - Clear Status 

This command deletes the status byte. It affects only the status byte of the interface, the command was sent from. No 
return value. For detailed description of the status byte see the different interface chapters. 

 

DCL - Device Clear 

This command resets the initialization data. No return value. 
Caution: Interface parameters are also reset! 

Содержание LAB-HP/E 10KW

Страница 1: ...ET System electronic GmbH Telefon 06205 3948 0 Fax 06205 37560 Hauptstra e 119 121 info et system de 68804 Altlu heim www et system de C LAB HP E...

Страница 2: ...Software version V15 November 2016...

Страница 3: ...Description of the different LEDs 9 Universal interface 10 Commands 10 Format 10 Instruction Set 10 Quick view of commands 10 Detailed description of commands 10 Response String 16 Ext Control Comput...

Страница 4: ...net 27 TELNET Connection with REALTERM 27 Ext Control USB 28 Interface Reconfiguration 28 Sense Mode 29 Load connection without Sensor Conductor 29 Load Connection without Sensor conductor 29 Load con...

Страница 5: ......

Страница 6: ......

Страница 7: ...ent manufacturing control through accredited laboratories and an ISO 9001 compliant quality management system guarantee a constant high level series production quality We offer repairs and adjustment...

Страница 8: ...pe D K 63A Type D K 63A Type D K 80A Type D K 120A Type D K 150A Type D K Leakage current cos phi Harmonic content 3 Efficiency Type 50Hz 72 100Hz 2 150Hz 0 9 200Hz 0 1 250Hz 11 350Hz 0 6 94 153 A 229...

Страница 9: ...e no Load Rel Accuracy 7 Voltage V 0 25 0 038 0 050 0 063 0 088 0 100 0 113 0 175 0 200 0 250 0 375 0 750 1 500 2 000 2 500 3 000 3 750 Current A 0 4 1 000 1 000 0 800 0 580 0 500 0 480 0 288 0 260 0...

Страница 10: ...4 ET System electronic GmbH IMPORTANT SAFETY INSTRUCTIONS...

Страница 11: ...oken connectors Do not use the device if any physical damage is apparent Please inform the carriers and a representative of ET System electronic immediately MAINS OPERATION Make sure to verify the mod...

Страница 12: ...Important Safety Instructions 6 ET System electronic GmbH...

Страница 13: ...fety Instructions ET System electronic GmbH 7 FUNCTIONAL DESCRIPTION THE FOLLOWING BLOCK DIAGRAM GIVES INFORMATION ABOUT THE VARIOUS ADJUSTMENT OPTIONS UI Mode Unit is oprated with current and voltage...

Страница 14: ...arameters for voltage current and OVP Over Voltage Protection Since the device is operated in UI mode only the set values for voltage and current will be transferred directly to the switching regulato...

Страница 15: ...pply to all interfaces OVP OVER VOLTAGE PROTECTION The output is shut immediately if the output voltage exceeds the selected value The display indicates this status with the word OVP To reset this err...

Страница 16: ...s the same GTL Gtl gTL Decimal places are optional and separated by a full stop The number of decimal places is not limited Therefore the effect of the following commands is the same UA 10 UA 10 0 UA...

Страница 17: ...AL button LIMI Limit Ia Reads maximum adjustable current limitation LIMU Limit Ua Reads maximum adjustable voltage limitation MI Nr Measure Ia Measures present output current MU Nr Measure Ua Measures...

Страница 18: ...This command adjusts current limiting Entering the command without parameters displays the set value If the set value is higher than the maximum current of the unit the range error bit within the ESR...

Страница 19: ...ery of maximum adjustable current LIMU 200 0V Unit answers Ulimit 200 0 V MI Nr Measure Ia This command measures the present output current Example LAB HP E 15100 100VDC 150A GTR Remote operation OVP...

Страница 20: ...e echo timeout Program Communication This command adjusts the interfaces The LAB HP E has a maximum of 3 digital interfaces x 1 2 or 3 The corresponding commands are PC1 PC2 or PC3 Type and number of...

Страница 21: ...bits 2 stop bits no handshake no parity echo on PC1 115200 N 8 2 N E Adjust baud rate to 115200 baud The new baud rate is active immediately after the command has been sent PC2 Query of second interfa...

Страница 22: ...without parameters displays the present set point If the set point exceeds the maximum voltage of the unit the range error bit within the ESR register of the interface is set The present set point rem...

Страница 23: ...U LIMU 500V UA UA 100V LIMI LIMI 30 00A OVP OVP 600V Example Command as ASCII and HEX protocol L I M U 5 0 0 0 V 4C 49 4D 55 2C 35 30 30 2E 30 56 0D 0A The digits after the decimal point correspond to...

Страница 24: ...ddress S1 S2 S3 S4 S5 Address S1 S2 S3 S4 S5 Address Off Off Off Off Off 0 Off Off Off Off On 16 On Off Off Off Off 1 On Off Off Off On 17 Off On Off Off Off 2 Off On Off Off On 18 On On Off Off Off 3...

Страница 25: ...the command STB or STB Return value STB xxxxxxxx Table Reading the status word Bit Function D7 n a D6 SRQ is set if SRQ was requested D5 ESB is set if a byte was set within the SES register D4 MAV is...

Страница 26: ...00 57600 62500 115200 Parity O Odd uneven parity E Even even parity N None no parity bit Number of data bits 7 or 8 Number of stop bits 1 or 2 Handshake H Hardware S Software N None no handshake The d...

Страница 27: ...erface sending the command PCx from one of the other interfaces using the display to configure the interface Interface Parameters RS485 INTERFACE OPTION The interface works with the following paramete...

Страница 28: ...op bits 0 1 Stop bit D4 Data format 1 8 bit 0 7 bit D3 n a D2 Table D1 Table D0 Table Table Error messages D2 D1 D0 Error 0 0 1 Syntax 0 1 0 Command 0 1 1 Range 1 0 0 Unit 1 0 1 Hardware 1 1 0 Read In...

Страница 29: ...tor output voltage Real on output 9 gnd GND 10 digital out Standby Signals standby 11 gnd GND 12 nc 13 REF 5 10 5 10 V Vref Output 5 10 V reference voltage 14 analog in Uset Set point U 15 analog in I...

Страница 30: ...h the word Error To reset this error standby mode must be activated Adjustment range is 0 V up to the maximum rated voltage of the device 20 Example LAB HP E with 600 V output voltage AI is adjusted t...

Страница 31: ...ivated by an applied voltage of 5 V up to 10 V Front panel operation is deactivated Operation mode is marked as AI on the display The digital interface takes priority over the AI interface The setting...

Страница 32: ...The status word can be read with the command STB or STB Only bits from D0 up to D2 are in use All other bits can be 1 or 0 Table Error messages D2 D1 D0 Error 0 0 1 Syntax 0 1 0 Command 0 1 1 Range 1...

Страница 33: ...trolling the device via port 10001 the user interface of the device must not be open in a browser TELNET CONNECTION WITH REALTERM Realterm is a free OpenSource terminal program which may be downloaded...

Страница 34: ...e RTS CTS D8 Software handshake XON XOFF D7 Parity enable D6 Parity mode 1 odd 0 even D5 Stop bit 1 2 Stop bits 0 1 Stop bit D4 Data format 1 8 bit 0 7 bit D3 used internally can be 1 or 0 D2 Table D1...

Страница 35: ...situation is essential Illustration 2 shows a correct load distribution Illustration 1 shows an insufficient supply of load 2 load 3 etc via parallel load conductors In practice it may occur that an...

Страница 36: ...evice resulting in a residual error of up to 0 5 depending on the load line used More detailed information can be obtained from the manufacturer WARNING INSTRUCTIONS FOR USING RELAYS TO THE LOAD DRAFT...

Страница 37: ...allel which doubles or triples the measured leakage current This is permissible according to the norm Quotation from the norm of 2008 appendix D When measuring protection conductor currents with the e...

Страница 38: ...asured ripple by using a bandwidth of 20MHz and the blue line shows the measured ripple by using a measure bandwidth of 300kHz This example shows that the measurement bandwidth have a strong influence...

Страница 39: ...a frequency analyser Bode 100 Also the used 1 1 Oscilloscope probe was during the tuning procedure connectet to the rippel measurment setup to elemenate the influence of the probe to the measurment b...

Страница 40: ...Notes 34 ET System electronic GmbH NOTES...

Страница 41: ...Notes ET System electronic GmbH 35...

Отзывы: