Measuring

8

5

 Once all items are verified, press the OK button to accept the images and return to  

  the Measuring screen.

  If only Standard and Position of Wear measurements are required, begin  
  Lens Boxing. For presbyopes, continue through the additional Personalization  
  Measurements before Lens Boxing.

Rotational Three-Quarter View

1

 Ask your patient to look at the bridge of his/her nose in the mirror.

Image Verification

3

 Press the Camera button to begin recording as the patient rotates  

  their head. Visioffice will record for a few seconds, automatically  
  selecting the most acceptable image with a Head Rotation angle  
  closest to 20°
.

4

 Verify three things:

  1. Your patient is focusing at his/her image in the mirror
  2. Your patient’s corneal reflections are visible and detected
  3. All markers are acurately aligned with the red targets

If these three conditions aren’t met, choose the image immediately before or after the automatic selection 
and apply the same checking process for corneal reflections and tracking markers.

Compare your patient’s Head Inclination from the Front View, verifying that the value variation does not 
exceed 3°. If the variation is greater than 3°, restart the two video captures with a more comfortable 
vertiwcal head posture in the Static Front View.

or

2

 While remaining focused on the bridge of their nose, ask the  

  patient to slowly rotate his/her head to the left or right.

Содержание Visioffice

Страница 1: ...Quick Reference Guide...

Страница 2: ...For Technical Assistance please contact the Visioffice Support Group at 1 877 457 3760 ext 3 or visioffice essilorusa com www essilorvisioffice com...

Страница 3: ...Contents Your Visioffice System Measuring Frame Styling Patient Education...

Страница 4: ...the UP position The Master Power Toggle is located in the rear access panel of the Visioffice Select the appropriate Patient Category from the Visioffice Home Screen based on the patient s prescripti...

Страница 5: ...nd ensure that the frame still rests against the back of the supports before returning the frame to your patient Patient Preparation The key to providing the best quality measurement results is proper...

Страница 6: ...r lower the Camera to patient s eye level 6 Position the Upper Frame Clip Marker on a horizontal line with the LE and RE references ensuring detection of all green trackers 7 Verify that the Frame Cli...

Страница 7: ...e same checking process for corneal reflections and marker detection Posture Verification Check the patient s Head Rotation and Head Inclination values remember them for later comparison Head Rotation...

Страница 8: ...ording Visioffice will record a video for a few seconds and select the most acceptable image Image Verification 3 Verify two things 1 Patient Corneal Reflections are visible and identified 2 All marke...

Страница 9: ...ead Rotation angle closest to 20 4 Verify three things 1 Your patient is focusing at his her image in the mirror 2 Your patient s corneal reflections are visible and detected 3 All markers are acurate...

Страница 10: ...ace in the gray area on the screen 4 Ask your patient to raise the Measuring Device with their arms outstretched focusing their vision through the open Alignment Zone and onto the Blue Optical Localiz...

Страница 11: ...et squarely facing the unit 3 Raise or lower the Visioffice camera to place your patient s face in the gray area on the screen 4 Ask your patient to read the text on the Handheld Measuring Device Pres...

Страница 12: ...to your patient s eye level ensuring detection of all four green clip markers If your patient is taller than the Visioffice arms adjust so the clips appear higher in the view If your patient is short...

Страница 13: ...vior is more naturally an Eye Mover If the Head Eye Ratio is 0 50 your patient s dynamic visual behavior is more naturally a Head Mover 2 Stability Coefficient A lower Stability Coefficient indicates...

Страница 14: ...erent patient before finalizing the current one you may do so since all pictures have been saved to the Patient Record Adjust the frame boundaries to the front face of the frame rim where the lens mee...

Страница 15: ...d designs requiring the Natural Head Posture measurement manually set your patient s head rotation Head Cape to the non zero value The measurement results will be adjusted automatically For non person...

Страница 16: ...re a still image or press the Video Camera button to record a short video clip 5 Once the images are captured you may display full screen or split view by using the numbered buttons on the left of the...

Страница 17: ...the screen 2 Use the tabs on the left of the screen to navigate through the each demo section Demos my be displayed in any order according to the needs of each patient Some demos can be customized aft...

Страница 18: ...Notes 17...

Страница 19: ...18 Notes...

Страница 20: ...lenses under the flagship Varilux Crizal DEFINITY Xperio Optifog and Essilor brands to correct myopia hyperopia presbyopia and astigmatism 2012 Essilor of America Inc All rights reserved Unless indica...

Отзывы: