background image

U

SER MANUAL

> S

AFETY CONSIDERATION

Nevertheless, special precautions need to be observed:

The use of accessories and cable other than those specified, with the exception of accessories and cables

sold of the slit lamp microscope as replacement parts for internal components, may result in increased

EMISSIONS or decreased IMMUNITY or decreased LIFESPAN of the slit lamp microscope.
The slit lamp microscope should not be used adjacent to or stacked with other equipment. In case

adjacent or stacked use is necessary, The slit lamp microscope should be observed to verify normal

operation in the configuration in which it will be used.

The length of cables or cords must be greater than 1.8 meters.

a. Electromagnetic emissions

This product is intended for use in the electromagnetic environment specified below. It is up to the

customer or the user to verify that the instrument is used in this environment.

E

MISSIONS TEST

C

OMPLIANCE

E

LECTROMAGNETIC ENVIRONMENT

– G

UIDELINES

Electromagnetic emissions (CISPR 11)

Group 1

The product uses RF energy only for its

internal function. Therefore, its RF emissions

are very low and are not likely to cause any

interference in nearby electronic equipment.

Disruptive voltage at power stations

(Conducted emissions) (CISPR 11)

Class A

The product is suitable for use in all

establishments other than domestic and those

directly connected to the public low-voltage

power supply network that supplies buildings

used for domestic purposes.

Harmonic current emission (IEC61000-3-2)

Class A

Voltage variations, voltage fluctuations and

flicker (IEC61000-3-3)

Complies

b. Magnetic and electromagnetic immunity

The product is intended for use in the electromagnetic environment specified below. It is up to the

customer or the user to verify that the instrument is used in this environment.

I

MMUNITY TEST

T

EST LEVEL

IEC

60601

C

OMPLIANCE LEVEL

E

LECTROMAGNETIC ENVIRONMENT

G

UIDELINES

Electrostatic discharge (ESD)

(IEC61000-4-2)

± 6 kV contact

± 8 kV air

± 6 kV contact

± 8 kV air

Floors should be wood, concrete or

ceramic tile. If floors are covered

with synthetic material, the

relative humidity should be at

least 30%.

Electrical fast transients and

bursts (IEC61000-4-4)

± 2 kV for power

supply lines

± 1 kV for the

signal ports

± 2 kV for power

supply lines

± 1 kV for the signal

ports

Mains power quality should be that

of a typical commercial or hospital

environment

Surge (IEC61000-4-5)

± 1 kV lines to

lines

± 2 kV lines to

earth

± 1 kV lines to lines

± 2 kV lines to earth

Mains power quality should be that

of a typical commercial or hospital

environment.

Power frequency (50/60Hz)

magnetic field

(IEC61000-4-8)

3 A/m

3 A/m

Power frequency magnetic fields

should be at levels characteristic

of a typical location in a typical

commercial or hospital

environment

< 5% U

T

< 5% U

T

Voltage dips, short interruptions

and voltage variations (IEC61000-

4-11)

(>95% dip in U

T

)

For 0.5 cycles
40% in U

T

(60% dip in U

T

)

For 5 cycles

(>95% dip in U

T

)

For 0.5 cycles
40% in U

T

(60% dip in U

T

)

For 5 cycles

Mains power quality should be that

of a typical commercial or hospital

environment. If the user of the

product requires continued

operation during power mains

interruptions, it is recommended

SL650L - Slit lamp microscope > V1 - 10-2021 - CE

33

Содержание SL 650

Страница 1: ...USER MANUAL...

Страница 2: ...a Intended purpose 9 b Indications for use 9 c Expected clinical benefit 9 d Intended population 9 e Intended users 9 IV INSTALLATION CONNECTION 12 a Checking program 17 a Turning OFF 19 V USE OF THE...

Страница 3: ...38 d Cleaning cycle 38 a Protection 39 b How to adjust the slit width of the control knob 40 c How to adjust the inclination of the illumination part 41 d How to replace the rail cover 41 e How to rep...

Страница 4: ...I INTRODUCTION...

Страница 5: ...and in any format whatsoever even free of charge is strictly prohibited without Mediworks s prior written consent This user manual is relevant for product SL650 It is based on relevant technical speci...

Страница 6: ...II SUPPLY PACKAGE...

Страница 7: ...he part of the device is presented before started the assembly 2 List of accessories While unpacking check that the following standard accessories are included Microscope Chinrest Binocular tubes part...

Страница 8: ...III GENERAL DESCRIPTION...

Страница 9: ...y inspection of contact lenses to judge the smoothness of the lens surface and the integrity of the lens 4 Assist orthokeratology lens fitting 5 Assist doctors in the examination and judgment of dry e...

Страница 10: ...or 7 Illumination system reset knob By loosening the knob the light can be offset from the center of the field of view of the microscope to provide indirect backlighting When the knob is tightened the...

Страница 11: ...e position 23 Fixation target Make the patient stare at it it is convenient for checking 24 Chinrest paper fixing bolt For fixing the pad paper 25 Chinrest Supporting the patient s chin 26 Focusing te...

Страница 12: ...IV INSTALLATION CONNECTION...

Страница 13: ...ake out the slit lamp part put it on the rails of the tabletop and check whether the wheels can be rolled steadily on the rails Place the rail cover to the rail remove four screws attached to the rail...

Страница 14: ...USER MANUAL INSTALLATION CONNECTION 1 Binocular tubes 2 Limit groove 1 Binocular tubes 2 Limit groove 1 Match here 2 Pin 14 SL650L Slit lamp microscope V1 10 2021 1...

Страница 15: ...ld directly on the gas shield screw 6 The 4 pin aviation plug of the chin rest is connected to the Output on the base of the slit lamp microscope the power adapter is connected to the Input on the bas...

Страница 16: ...N 7 The power switch can be turned on with one short press short pressed again to hold and long press for 3seconds to turn off all power The power supply can work normally from 100V to 240V SL650L Sli...

Страница 17: ...ness change of the illumination should be observed when rotate the brightness adjustment knob 3 Check the fixation target to verify that it is properly illuminated 4 Check whether the moving parts suc...

Страница 18: ...ded well 1 A 3 pin cable is supplied with this instrument Correct plug is supplied as well Ensure the instrument is grounded 2 The indicator lamp will be lighted when the instrument is power ON 3 Inse...

Страница 19: ...ht 7 Turn OFF the main power and cover the instrument with the dust proof cover after testing a Turning OFF Turn OFF the main power and cover the instrument with the dust proof cover after testing 3 C...

Страница 20: ...V USE OF THE DEVICE...

Страница 21: ...operator side 1 The flat faces microscope 2 Insert the test rod into the hole 2 Brightness adjustment Turn on the main power switch and turn the brightness adjustment knob to the middle Adjust the sli...

Страница 22: ...image When adjusting the pupil distance ensure that the two eyepieces are at the same height Tilt the binocular tubes base on both sides 2 Patient s head position and use of fixation target 1 Place t...

Страница 23: ...a clear observation 1 Joystick 2 Clockwise to lower counterclockwise to raise 4 Lock the base After the microscope is adjusted tighten the base locking screw to fix the base so that it cannot move 4 O...

Страница 24: ...can be displayed by the scale board Each small grid is 5 and the larger grid is 10 4 Offset illumination Rotate the centering knob and rotate the slit width adjustment knob in the direction of the ar...

Страница 25: ...the position of the heat absorbing filter after use 7 From left to right white filter heat absorbing filter light reduction filter red free filter cobalt blue filter The white film is only used when...

Страница 26: ...lamp microscope It is necessary for proper operation and diagnosis 2 During the use of the operator the different scales and different indication marks corresponding to the various knobs at different...

Страница 27: ...USER MANUAL USE OF THE DEVICE 26 SL650L Slit lamp microscope V1 10 2021 CE...

Страница 28: ...VI ERROR DISPLAY...

Страница 29: ...USER MANUAL ERROR DISPLAY This section is not applicable 28 SL650L Slit lamp microscope V1 10 2021 CE...

Страница 30: ...VII SAFETY CONSIDERATION...

Страница 31: ...us injury Danger a hazardous situation that if not avoided will result in death or serious injury Important and or useful additional information to learn relating to the text in this manual Tips pract...

Страница 32: ...ional maintenance personnel to help you troubleshoot 3 Do not store and use in a flammable explosive high temperature high humidity and dusty environment use it in a clean room keeping the product cle...

Страница 33: ...responsible for damage caused by fire earthquake third party behavior other accidents and carelessness of the user misuse or use under abnormal conditions Manufacturer is not responsible for the defic...

Страница 34: ...0 3 2 Class A Voltage variations voltage fluctuations and flicker IEC61000 3 3 Complies b Magnetic and electromagnetic immunity The product is intended for use in the electromagnetic environment speci...

Страница 35: ...mitter Recommended separation distance d 1 167 P d 1 167 P 80 MHz to 800 80 MHz to 800 MHz d 2 333 P 800 MHz to 2 5 GHz Where p is the maximum output power rating of the transmitter in watts W accordi...

Страница 36: ...VIII TROUBLESHOOTING...

Страница 37: ...knob is at minimum Set the brightness adjustment knob Slit is too dark Voltage selector is wrongly set Set the voltage selector correctly The coat of the reflecting mirror is oxidized Change the refl...

Страница 38: ...IX MAINTENANCE...

Страница 39: ...d in absolute alcohol Don t wipe with hands or hard object or any corrosive detergent lest that the surface should be damaged b Sliding pad rails and shaft Clean these parts with clean soft cloth regu...

Страница 40: ...uld be replaced when inspecting each of the inspectors The headrest strap area should also be cleaned Both parts should be cleaned and disinfected daily before the first use 4 For the whole device Cyc...

Страница 41: ...the locking screw on right side clockwise with a hexagon type screw driver 1 Left knob 2 Right knob 3 Locking screw on the right knob 2 Rotate the locking screw with a hexagon type screw driver If The...

Страница 42: ...tion mechanism of the illumination part is too loose fasten the screw on both sides of the pivot point with the screw driver Screw d How to replace the rail cover 1 Place your hand on the rail cover 2...

Страница 43: ...USER MANUAL MAINTENANCE 3 4 Don t Move the rail cover in the direction of the arrow shown If the mid cover area is detached the disassembly is successful 42 SL650L Slit lamp microscope V1 10 2021 CE...

Страница 44: ...es The pictogram of the wheeled container appears on the label of the instrument It indicates the obligation for separate collection and disposal of end of life out of use electrical and electronic eq...

Страница 45: ...X SPECIFICATIONS...

Страница 46: ...length Continuously variable from 1 mm to 14 mm Aperture diameters 14 mm 10 mm 5 mm 3 mm 2 mm 1 mm 0 2 mmSlit angle 0 180 Adjustable both vertical and horizontal Slit inclination 4 steps 5 10 15 20 Fi...

Страница 47: ...USER MANUAL SPECIFICATIONS 45...

Страница 48: ...USER MANUAL SPECIFICATIONS 2 It requirements This section is not applicable 3 Connectivity to other devices This section is not applicable SL650L Slit lamp microscope V1 10 2021 CE 46...

Страница 49: ...XI QR CODE...

Страница 50: ...USER MANUAL QR CODE User manual in the appropriate language is available on a web space Upon request a paper version can be provided for free 48 SL650L Slit lamp microscope V1 10 2021 CE...

Страница 51: ...USER MANUAL QR CODE SL650L Slit lamp microscope V1 10 2021 CE 49...

Страница 52: ...Essilor Instruments USA 8600 W Catalpa Avenue Suite 703 Chicago IL 60656 Phone 855 393 4647 Email info essilorinstrumentsusa com www essilorinstrumentsusa com...

Отзывы: