Bench-top type temperature(humidity) chamber
Installation guide
20
ESPEC CORP.
4) Tighten the water tank cap and set the tank back in the water circuit
compartment.
* Slide the tank inward until hearing the connection plug catch.
When the connection plug locks onto the water circuit, the plug's valve
opens and water runs freely into the circuit.
Gently try pulling the tank outward. If the tank comes out, it is not
connected. Set the tank back in the water circuit compartment.
◆
Note
◆
Do not run temperature and humidity operations with the water tank out. Pull
the tank out only to add to deionized water to it. In this state, water is not
supplied to the humidifying tray or wick pan, which can prevent temperature
and humidity from attaining or keeping the set targets.
Do not duplicate setting water tank to water circuit compartment for many
times. It may create a loss of pressure balance. As the result, siphon effect
can occur in water tank level sensor tank due to loss of pressure balance, it
may invite water is drained continuously to outside chamber.
CAUTION
Be sure to close the tank front cover except when the tank is pulled
out.
It may result in injuries by the front cover when you put your hands into
the water circuit compartment.
To supply water to the humidifying tray
To supply water from the water supply tank to the humidifying tray, it is
necessary to run the chamber. The chamber is set to 0°C and humidity control is
OFF when the chamber is shipped from the factory.
< Procedure >
1) Turn the circuit breaker ON.(On the right wall of the chamber)
2) Press the
POWER switch on the instrumentation.
3) The monitor screen will appear.