Operation
72
Platinous User’s Manual
Droop cables below and up to specimens as shown in
If cables are taut, dew can run down the cables and damage specimens.
To charge specimens with less than 2A of power
Be sure to fit the line with a suitable capacity fuse.
FIGURE 6.44.
W
IRING
TO
SPECIMENS
(
LESS
THAN
2A)
To charge specimens with 2A or more power
Be sure to fit the line with a suitable capacity contactor and fuse.
FIGURE 6.45.
W
IRING
TO
SPECIMENS
(2A
OR
MORE
)
2. Turn the time signal ON for Steps 2 and 4 of the program. See
6.5.3 External Alarm Terminal
Purpose - Use
This feature is used to notify an external point when the chamber has been stopped by a safety device
tripping.
!
CAUTION
S p e c im e n
T e s t a re a
P o w e r s u p p ly fo r
s p e c im e n s
F u s e
S p e c im e n p o w e r
S u p p ly c o n tro l
te rm in a ls
T im e s ig n a l
te rm in a ls
F ig . 5 .1 5 W irin g to S p e c im e n s
(T o c h a rg e s p e c im e n s w ith le s s th a n 3 A o f p o w e r)
Specimen
Contactor
Fuse
Fuse
Power supply for
specimens
Test Area
Specimen power
Supply control
terminals
Time signal
terminals
Содержание Platinous ESL Series
Страница 1: ...PLATINOUS CHAMBER SERIES USER S MANUAL...
Страница 2: ......
Страница 6: ...iv Platinous User s Manual...
Страница 22: ...Precautions in Usage 16 Platinous User s Manual...
Страница 31: ...Platinous User s Manual 25 Names and Functions of Parts...
Страница 40: ...Names and Functions of Parts 34 Platinous User s Manual...
Страница 58: ...Installation 52 Platinous User s Manual...
Страница 80: ...Operation 74 Platinous User s Manual...
Страница 106: ...Troubleshooting 100 Platinous User s Manual...
Страница 107: ...Platinous User s Manual Chapter 9 Appendix...