Installation
46
Platinous User’s Manual
5.7 Water Level Check (ESL, ESX models only)
For proper chamber operation, water level in the humidity tray must be checked.
5.7.1 Remove Protection Material
The wick pan water level regulator and the humidifying tray water level regulator are contained inside the
water circuit compartment, located on the right side of the chamber.
Do not re-install float upside down!
Procedure
1. Open the lids of both the water level regulators.
2. Take out the stopper sponges (protection material) found inside.
3. Note the position of the floats when the sponges are removed.
4. Return and close the lids to their original state.
FIGURE 5.26.
W
ATER
LEVEL
REGULATORS
The valve seat lock nut, fitted on the lid of the wick pan water level
regulator, and the float switch holder lock nut fitted on the
humidifying tray water level regulator are factory preset and should
not be turned.
The float for the wick pan water level regulator and the float for the
humidifying tray will not function properly if interchanged.
When re-adjusting either of the level regulators, refer to the instructions on
Water Level in the Humidifying Tray Water Level Regulator” on page 49
Water Level in the Wick Pan” on page 50
!
CAUTION
Valve Seat
Locknut
Lid
Sponge
Float
Float Switch
Lock Nut
Lid
Sponge
Float
Wick Pan Water Level Regulator
Humidifying Tray Water Level Regulator
!
CAUTION
Содержание Platinous ESL Series
Страница 1: ...PLATINOUS CHAMBER SERIES USER S MANUAL...
Страница 2: ......
Страница 6: ...iv Platinous User s Manual...
Страница 22: ...Precautions in Usage 16 Platinous User s Manual...
Страница 31: ...Platinous User s Manual 25 Names and Functions of Parts...
Страница 40: ...Names and Functions of Parts 34 Platinous User s Manual...
Страница 58: ...Installation 52 Platinous User s Manual...
Страница 80: ...Operation 74 Platinous User s Manual...
Страница 106: ...Troubleshooting 100 Platinous User s Manual...
Страница 107: ...Platinous User s Manual Chapter 9 Appendix...