Esma E188T Скачать руководство пользователя страница 5

 

1.

 

Heater – Tank is supplied with 450 watts of heat controlled by an    
adjustable, internal thermo-switch. The thermo-switch has been    
adjusted to control the bath temperature at 120degrees F when tank     
covers are on. If temperature needs to be changed: 

o

 

Disconnect unit from 120Voutlet. 

 

o

 

Remove back panel. 

o

 

Move control knob of thermostat slightly in clockwise 
direction to increase temperature and counter-clockwise to 
decrease temperature. Make only slight changes because a 
30degree angle change corresponds to an approximate 
20degrees F temperature change. Replace top plate on 
chamber. Use caution that the bath temperature does not 
exceed 180degrees F. 

 

 

 
 
 
 
2.

 

Timer – The unit has been equipped with an Esma digital timer.  The 
timer is preset at 15 minutes.  To change the time simply press the 

 

or 

 buttons accordingly.  Press the Start/Stop button to activate the 

timer.  Once the timer is running, pressing the Start/Stop button will 
stop the ultrasonics.  Pressing the Start/Stop button again will 
resume the ultrasonics. 

 
 
 

 
 
 

3.

 

Tumbler 

 

Timer

 position- ON only with timer. 

 

Continuous

 position- ON with main switch. 

 
 
 

Отзывы: