Celebrating over 30 years of excellence in manufacturing ultrasonic cleaning
equipment, electropolishing equipment and associated chemistries
ESMA, Inc.
P. O. BOX 734 * SOUTH HOLLAND, IL 60473 * (800)-276-2466 * FAX (708)331-8919
7
4. ELECTRICAL
The unit is rated at 1500 W, 120 VAC, 50/60 HZ. Only the power module is plugged into your 120VAC supply (diagram).
The Tank Module has two plugs.
Both plugs are attached to power module at receptacles C & E.
Control Module
plug (M) plugs into power module receptacle (G).
Only the plug from power module
(A) is to be connected to your power source. Unit must be electrically grounded. The
power cord must be connected to a three-way grounded outlet. For 2-wire service, an adapter with external ground wire is
necessary. Connect the green grounding wire of the adapter to the screw, which holds the electric outlet plate cover to the
socket. DO NOT OPERATE UNIT WITHOUT PROPER GROUNDING. A fuse, 15AMP, 120VAC, is located on the power
module.
5. CONTROL FUNCTION
A. Main Switch:
(Located on Power Module) When unit is ready for operation, put the main
switch to ON and the
indicator light on the Power Module and the Control Module will be
ON
.
B. Run-Reset Switch:
Located on Control Module, must be in the RUN position for the programmable controller to
operate. If during the program cycle you want to stop the process, turn switch to RESET position and the program
cycle will discontinue. Move switch back to RUN position and program will START over at the beginning of the fill
cycle when start button is pushed.
Use run-reset switch only for emergency stopping of unit.
C. Fill/Clean/Full Cycle Switch:
Fill and Hold Cycle:
Must start at FILL position and push start button. After tank is filled the cycle will stop
and allow instruments to soak prior to continuing the clean/rinse/dry steps. Move switch to CLEAN position. When
instruments or parts have been added, push START button and the clean/rinse/dry cycle will begin. You cannot