EsiWelma Sensigas URD20SS Скачать руководство пользователя страница 4

EsiWelma Srl 

URD20SS_en.doc - rev. 0 

CO

2

 gas detectors – URD20SS 

24/09/2010 

Gas detection systems for industrial environments 

4/8

 

Electrical Installation 

 
 

 
 
 
 
 
 

Terminal board and 
electrical connections 
 
 
 
 
 
 
 
 
 

 
 

Cabling:

 

 
 
 

 

Configuration: 

 
 
 
4…20mA Output 
reference selection: 
 
 
 
 
 
 
4...20mA signal 
operating mode 
configuration: 
 
 

 

Setting threshold limit 
values: 
 
 

CAUTION:  

 Make the area safe and make sure that the device power supply is 

off before cabling and configuration operations.

 

Install the sensor in compliance with EN 60079-14. 
The cable gland provided on the housing is used for cable entry. 
Ground the sensor using the internal grounding system. 
 

Ground connection 
 
 
Terminal Board TB1 

+

12…24Vdc

 


4…20mA

 

+

 
JP2 jumper circuit 

 

S1 DIP switch 

 

 

 

 
 

 
 

Depending on the connecting distance, use at least 3-core cable, min diameter 
0.75mm

2

 up to 100mm

2

, 1mm

2

 up to 200mm

2

 , 1.5mm

2

 up to 500mm

2

 . 

Use shielded cable where there is a risk of electromagnetic interference.  
If a relay card is used, use multi-core cable suitable for the number of connections. 

 

Default settings of the sensor are shown in the “Technical Specifications” chapter. 
In order to change the default settings, switch off the power supply, input the new 
settings at the JP2 jumper circuit or at the S1 DIP switch as shown in the diagram, 
then power-up again; in particular: 

 

The default setting for the 4-20mA signal is the negative power signal. Output 
reference selection should be made by JP2 triple of jumpers; to change this setting, 
operator has to move 

JP2

 jumpers as shown in the figure: 

Caution:

 if the default settings are changed, 

the connections on the 

TB1

 terminal board 

will be inverted.

 

     

     Ref. at  -  (default)            Ref. at    

 

 

To set the operating mode of the 4...20mA signal, it is necessary to use the 5th 
selector of the DIP switch at 

S1

; in particular: 

 
 

 

                                            Proportional                        Threshold mode 

 

To set the threshold limit values of the optional relay card, or of the threshold 
operating mode of the 4...20mA signal, it is necessary to use the first four selectors 
of the DIP switch at 

S1

; in particular, the thresholds, given in full scale percentage, 

will be: 
 
 
 
 
 
 
 
 
 
 
 

 

Sensor 
card 

(*) When the first four selectors
of the DIP switch are in OFF
position, the threshold limit
values can only be set by the

TUS40

 handheld terminal. 

If this is selected without using
the handheld terminal, the
detector will automatically set
the default threshold limit
values.    To set the detector
with the handheld terminal, see
the dedicated instruction
manual. 

Relay card 

1   2   3   4   5   6   7   8 

ON 

1   2   3   4   5   6   7   8 

ON 

CN4 connection for handheld 
terminal 

Connection slot 
for relay card 

CUSTOM (*)              3, 5, 10%               5, 10, 15%             5, 10, 20%             10, 15, 25%           10, 15, 30% 

10, 20, 40%             10, 25, 35%            15, 25, 40%            15, 30, 45%            25, 35, 50%           20, 40, 60% 

(DEFAULT) 

20, 40, 80%            -----------------Future implementations------------------ 

1   2   3   4   5   6   7   8 

ON 

1   2   3   4   5   6   7   8 

ON 

1   2   3   4   5   6   7   8 

ON 

1   2   3   4   5   6   7   8 

ON 

1   2   3   4   5   6   7   8 

CU

1   2   3   4   5   6   7   8 

ON 

1   2   3   4   5   6   7   8 

ON 

1   2   3   4   5   6   7   8 

ON 

1   2   3   4   5   6   7   8 

ON 

1   2   3   4   5   6   7   8 

ON 

1   2   3   4   5   6   7   8 

ON 

1   2   3   4   5   6   7   8 

ON 

1   2   3   4   5   6   7   8 

ON 

1   2   3   4   5   6   7   8 

ON 

1   2   3   4   5   6   7   8 

ON 

1   2   3   4   5   6   7   8 

ON 

Содержание Sensigas URD20SS

Страница 1: ...etects carbon dioxide leaks or emissions in industrial environments hospitals fermentation plants greenhouses stables and more in general where carbon dioxide is stored generated or produced URD20SS sensors can be used in stand alone mode with 4 20mA output or with an optional voltage free contact relay card having the following four digital outputs Pre alarm 1st alarm threshold 2nd alarm threshol...

Страница 2: ...larm by jumper selectable polarity 4 20mA output load resistor up to 200Ω at 12Vdc power supply 200Ω 700Ω at 24Vdc power supply Operating Temperature Storage Temperature 20 50 C 20 70 C Relative Humidity without condensing Operation Storage 15 90 RH 45 75 RH Operating pressure KPa 80 110 Air speed m s 6 Visual warnings Red LED visible with detector energised Dimensions and weight See dedicated sec...

Страница 3: ...s that of air so it tends to collect at floor level in closed unventilated environments Therefore the sensor must be installed about 30 cm above the floor level Take into consideration the following specific installation guidelines as well as the above instructions for location of the detectors The detectors must be installed 1 where accidental gas leakages are possible 2 at least 1 5m away from h...

Страница 4: ...be made by JP2 triple of jumpers to change this setting operator has to move JP2 jumpers as shown in the figure Caution if the default settings are changed the connections on the TB1 terminal board will be inverted Ref at default Ref at To set the operating mode of the 4 20mA signal it is necessary to use the 5th selector of the DIP switch at S1 in particular Proportional Threshold mode To set the...

Страница 5: ...s below Step 1 Insert the connection slot provided with the relay card into the control card making sure the flexible tab is towards the main terminal board Find CN3 connector CN3 connector Connection slot flexible tab Step 2 Fit the relay card snugly and pull the flexible tab of the connection slot towards the main terminal terminal board Flexible tab Step 3 Check the position of the card Make su...

Страница 6: ...al installation The sensors are factory calibrated so they normally do not require any other calibration once installed Still after installation an operational check of the sensors is recommended The detector will enter a 2 minute warm up phase after power up After this time the sensor will switch to normal operating mode but it will take about 2 hours before it reaches top performance level When ...

Страница 7: ...to the supplier installer who in turn will send it back to the manufacturer Sensors may need to be re calibrated using the TUL40 gas calibration kit and the TUS40 handheld terminal which must be connected to the sensor via the communication interface integrated in the cable on the connector CN4 For the re calibration procedure see the instructions supplied with the handheld terminal Remove power f...

Страница 8: ...use in explosive gas atmosphere caused by mixture of air and gas vapours flammable mist Ex nA nC d IIC T6 Types of protection complying with EN60079 0 and EN60079 15 sensor body with type of protection d in compliance with EN60079 1 BVI 07 ATEX 0033 Type examination certificate 20 C TA 50 C Operating temperature range Installation data To be filled in by Installer Installer s stamp and signature I...

Отзывы: