background image

Portable Ionization Test Kit
Operation and Maintenance

PS-2062

6/05 Page 1 of 4

Tech Brief

Made in America

© 2005 DESCO INDUSTRIES INC.

Employee Owned

Description

Our Portable Ionization Test Kit is a
portable and cost effective means of
verifying the performance of a wide
variety of ionization equipment. The
Test Kit includes a 42721 Digital
Fieldmeter, an Isolated Plate
Assembly, and a Charger. The
Charger is used to place a 1000V
charge on the isolated adapter plate
assembly, making it possible to also
measure the neutralization
performance of air ionization
equipment. 

NOTE:

Our 42721 Digital

Fieldmeter is designed to operate
with our 42755 Portable Ionization
Test Kit. It is not compatible with
other brands.

Currently there are no established
standards describing a periodic
verification device or procedure for
air ionizers. The ESD Systems.com

Portable Ionization Test Kit has been
designed to make measurements that
correspond to those made by using a
charged plate analyzer and ESD
Association's Standard 3.1. While
the Ionization Test Kit provides
convenience and portability, it does
not meet all of the requirements of
the ESD Association Standard. ESD
Systems.com recommends our
model 42630 Charged Plate
Analyzer if precise measurements
are required.

The Portable Ionization Test Kit
includes a slide-on isolated plate
assembly, a ±1000 volt charging unit
and a durable thermoplastic carrying
case with custom cut-outs for all of
the above components along with
the model 42721 fieldmeter. The
42721 Digital Fieldmeter can be
purchased separately.

Inspection

Remove the kit from the carton and
inspect for damage. Each kit
includes:

1 Digital Fieldmeter (Item 42721)
1 Isolated Plate Assembly
1 ± 1000 Volt Charger
1 9 Volt alkaline battery (installed)
1 Carrying case

Taking Balance
Measurements

The Portable Ionization Test Kit has
been designed to match the compact
size and hand held convenience of
the 42721 Digital Fieldmeter. Use
the following procedure to verify the
balance of air ionization equipment.
This quick and easy procedure will
indicate if the piece of ionization
equipment is working within the
manufacturer's specifications or user
requirements. 

It is extremely

important that ionizers be checked
regularly for balance. An ionizer
operating in an out-of-balance state
can damage sensitive electronic
components or assemblies.

1. INSTALLING THE ISOLATED
PLATE ASSEMBLY -

The case of

the Model 42721 Fieldmeter has two
slots along its slides. The top slot is
closest to the front of the instrument.
Slide the tabs of the Isolated Plate
Assembly into the top slot of the
meter case as far as they go.

Figure 1.  42755 Portable Ionization Test Kit

Содержание 42755

Страница 1: ... precise measurements are required The Portable Ionization Test Kit includes a slide on isolated plate assembly a 1000 volt charging unit and a durable thermoplastic carrying case with custom cut outs for all of the above components along with the model 42721 fieldmeter The 42721 Digital Fieldmeter can be purchased separately Inspection Remove the kit from the carton and inspect for damage Each ki...

Страница 2: ...R is disabled the decimal point will be on continuously Holding down the ZERO button while unit is ON for less than 3 seconds resets the AUTO SHUTDOWN TIMER if enabled The AUTO SHUTDOWN TIMER is enabled or disabled by turning on the unit and keeping the ON ZERO button pressed then toggling the HOLD OFF button Enable disable of the AUTO SHUTDOWN TIMER is indicated by the decimal point DP on timer o...

Страница 3: ...he Charger in one hand and the Fieldmeter with Adapter Plate in the other If this is not possible test leads must be used to connect one to the Charger output terminals and the Fieldmeter ground connection Calibration The overall accuracy of the 42755 Test Kit is 5 nominal Calibration of the Ionization Test Kit should only be performed by the factory The 42755 is calibrated to NIST traceable stand...

Страница 4: ...charge Call Customer Service at 508 485 7390 for a Return Material Authorization RMA and proper shipping instructions You should include a copy of your original packing slip invoice or other proof of purchase date Any unit under warranty should be shipped prepaid to the ESD Systems com factory Warranty repairs will take approximately two weeks If your unit is out of warranty ESD Systems com will q...