CelCulture
6
2.1.3 Power Requirements
The equipment is designed to work with an electrical supply of 220 VAC – 240 VAC, 50/60 Hz or
110VAC – 130 VAC, 50/60 Hz. If your available electrical supply is not within these parameters, then a
suitable power supply must be used, otherwise damage may be caused to the device or a hazardous
situation may result.
It is recommended that the equipment is connected to a dedicated power source with protective
grounding installed
There should be unobstructed access to the main power plug. The power plug is the main
disconnecting device on the unit
This equipment can sustain a maximum of ±10% nominal voltage fluctuation; otherwise a power
stabilizer is needed.
2.2 Installation
2.2.1 Unit Leveling
It is important that the equipment is properly leveled by positioning a bubble level on the center top shelf of
the incubator. The incubator has a built in leveling feet that can be adjusted to level the unit. Adjust the feet
until the unit sits level left to right and front to back.
2.2.2 Unit Stacking
If equipment is to be stacked on top of each other,
the equipment with the lower working temperature
must always be placed at the bottom.
1.
Remove the screws from the back at top
left and top right of the incubator. These
screws will be used to secure the stacking
brackets later.
2.
Position a bubble level on the center top
shelf of the incubator and adjust the
leveling feet of the lower incubator until
the incubator is level.
3.
Lift the 2nd incubator onto the top the 1st
incubator and align them.
Note:
The feet are NOT to be removed and they
should be adjusted to allow 40 mm to 50 mm
(1.5” to 2”) clearance between the incubators.
4.
There are 4 pieces provided in the
accessory kit box, 2 for the front and 2 for
the back.
5.
Secure the stacking brackets to the cabinet
using the M4 screw provided.
6.
Tighten the nut on the foot to secure the
stacking brackets.
2.2.3 ULPA Filter Installation
To install the ULPA filter (2), push it upward in place – just beneath the top
duct (1). To uninstall, pull it downward.
1
2