28
Problem
1:
Cabinet
does
not
start
(LCD,
button,
fan
light,
and
socket
are
inoperative)
Cause
Corrective
Action
Power
Failure
•
Check
if
there
is
power
at
the
wall/building
electrical
socket
by
using
DVM.
•
Ensure
the
building
socket
switch
is
at
ON
position.
Power
cord
is
not
connected
properly
or
faulty
•
Check
whether
power
cord
has
been
connected
properly
into
wall/building
socket
and
the
unit.
Some
cabinets
have
2
cords.
•
Measure
the
AC
voltage
between
the
live
and
the
neutral
terminal
of
the
cord
by
using
DVM.
•
If
the
voltage
is
not
within
±2%
of
the
wall
socket
voltage,
replace
cord;
otherwise
proceed
to
next
step
Circuit
breaker
has
tripped
•
See
Component
Layout
to
find
circuit
breaker.
•
Check
the
circuit
breaker
inside
the
electrical
panel.
NOTE:
If
circuit
breaker
has
tripped,
do
not
reset
the
breaker
before
checking
all
electrical
components
and
wiring
connections.
(See
Figure
1
‐
1).
•
Does
the
cabinet
operate
correctly
after
resetting
the
circuit
breaker?
If
not
proceed
to
next
step.
Improper
connection
•
Ensure
cabinet
is
plugged
in
to
the
main
supply.
•
Carefully
measure
AC
voltage
between
LIVE
and
NEUTRAL
terminal
block
inside
electrical
box
(See
Figure
1
‐
2).
•
The
voltage
should
be
230VAC
±10%
for
ADC
‐
__1/3
or
115VAC
±10%
for
ADC
‐
__2.
•
If
the
voltage
is
out
of
the
range,
check
cable
connection
at
connector
A
pin
1
(live)
,
2
(neutral)
and
3
(ground).
Confirm
that
the
voltage
is
present
•
See
Component
Layout
to
locate
connector
A
and
the
pin
numbering
system.
Figure
1
‐
1
Figure
1
‐
3
Figure
1
‐
2
Содержание Ascent max
Страница 6: ...iv...
Страница 10: ...viii...
Страница 22: ...12...
Страница 26: ...16...
Страница 60: ...50...
Страница 61: ...51 Ductless Fume Hoods APPENDIX...
Страница 70: ......