1.2.3 Technical specification summary table
General Specifications, Aeris Thermal Cycler
Block Code
AERIS-BG086
AERIS-B4830
AERIS-
BG384
AERIS-BG384
AERIS-B4076
Sample Capacity
96 × 0.2ml
48 × 0.2ml + 30
× 0.5ml
384 wells
48 × 0.2ml + 48
× 0.2ml
4 slides
Applied
Consumables
0.2ml tubes
96 micro plate
12 x 8 strips
8 x 12 strips
0.2 ml tubes
0.5 ml tubes
4 x 12 strips
384 micro
plate
0.2ml tubes
6 x 8 strips
4 slides in situ
plate
Temperature
Range
4ºC ~ 105ºC
Max. Heating
Rate
4.0ºC / sec
2.8ºC / sec
2.8ºC / sec
4.0ºC / sec
1.8ºC / sec
Max. Cooling
Rate
4.0ºC / sec
2.8ºC / sec
2.8ºC / sec
4.0ºC / sec
1.8ºC / sec
Temperature
Uniformity
≤
0.2ºC
Temperature
Accuracy
≤
0.1ºC below 50 ºC
Max. Gradient
30ºC
-
30ºC
-
-
Min. Gradient
1ºC
-
1ºC
-
-
Hot Lid Temp.
range
30ºC ~ 110ºC
Temp. Control
Mode
Block mode or Tube mode
Display
6.5” Color LCD Touch Screen
Instrument
Memory
Up to 250 programs, unlimited with USB flash drive
PC Interface
RJ45 & USB port
Dimensions
(W x D x H)
306mm x 386mm x 295mm (WxLxH)
Weight (kg)
10
Power Supply
100-240 V / 50-60 Hz 600W
Fuse
250V 8A ø5 × 20
Warranty
3 years on main body, 2 years on blocks
1.3
Security Notes
i. Turn the power OFF and unplug the power cord before performing any procedure.
ii. Please read all the instructions thoroughly before maintenance. Always follow this
service manual. Keep this service manual well for future reference and review it as
necessary.
iii. Engineers must inspect and ensure operation of the equipment is completely back to
normal after maintenance.