Step 5:
Connect the inflator
Outlet
to a second stable >100 PSI air source.
Step 6:
The current air pressure will be shown on the inflator’s LCD display.
Step 7:
Once the air pressure reading is steady and holding its position press the “
i/SET
” unit
option button once on the inflator.
The display screen will show a reading of “
100 PSI
”, at this point shut off the air source
and calibration is complete.