32
EMGen Command Set
ets-lindgren.com
AM:DEPT:MAX?
Description:
Gets the maximum amplitude modulation depth as a %.
Syntax:
[:SOURce]:AM[:DEPTh]:MAXimum?
Example:
Write:
AM:DEPT:MAX?
Response:
:AM:DEPT:MAX 100.0
AM:DEPT:MIN?
Description:
Gets the minimum amplitude modulation depth as a %.
Syntax:
[:SOURce]:AM[:DEPTh]:MINimum?
Example:
Write:
AM:DEPT:MIN?
Response:
:AM:DEPT:MIN 0.0
AM:INT:FREQ
Description:
Sets the amplitude modulation frequency in Hz. The units default to Hz but can
also be specified as KHZ, MHZ or GHZ.
The value can also be increased or decreased by step size:
UP = Increase amplitude modulation frequency by step size.
DOWN = Decrease amplitude modulation frequency by step size.
Use the AM:INT:FREQ:STEP command to change the current step size.
Syntax:
[:SOURce]:AM:INTernal:FREQuency <value><unit>|UP|DOWN
Example:
Write:
AM:INT:FREQ 1000
Write:
AM:INT:FREQ 2 KHZ
AM:INT:FREQ?
Description:
Gets the amplitude modulation frequency in Hz.
Syntax:
[:SOURce]:AM:INTernal:FREQuency?
Example:
Write:
AM:INT:FREQ?
Response:
:AM:INT:FREQ 2000