16
Operation
www.ets-lindgren.com
Probe Selection
Choosing the correct probe is determined by the following:
•
Whether the signal is E or H:
If the signal is primarily is E-field, use
the ball probe or stub probe. If the signal is primarily H-field, use one of
the loop probes. If unknown, try one of each and select the one that
best picks up the signal.
•
The strength of the signal:
Select a probe that adequately receives
the desired signal of interest. Respectively, the ball probe and the
6-cm loop are the most sensitive of the E-field and H-field probes. The
stub probe and the 1-cm loop are the least sensitive.
•
The frequency of the signal:
If the signal is above 790 MHz, the
probe may go into resonance. See the upper resonant frequency listed
for each probe in
Specifications
on page 13.
In this illustration a
ball probe is used to
examine a flat cable.
The distributed
inductance over the
length of the cables
makes them
particularly susceptible
to common mode
problems. High
impedance sources
such as this are best
examined with an
E-field probe.
•
The physical size of the space where the probe must fit:
Model 7405 includes a variety of sizes. See pages 8–9 for a
description of each probe.
Содержание ETS-Lindgren 7405
Страница 1: ...Model 7405 Near Field Probe Set User Manual...
Страница 2: ...ii www ets lindgren com...
Страница 12: ...12 Maintenance www ets lindgren com This page intentionally left blank...
Страница 18: ...18 Operation www ets lindgren com This page intentionally left blank...
Страница 20: ...20 Typical Performance Factors www ets lindgren com Magnetic H Field Probes 901 6 CM LOOP...
Страница 21: ...www ets lindgren com Typical Performance Factors 21 902 3 CM LOOP...
Страница 22: ...22 Typical Performance Factors www ets lindgren com 903 1 CM LOOP...
Страница 23: ...www ets lindgren com Typical Performance Factors 23 Electric E Field Probes 904 BALL PROBE...
Страница 24: ...24 Typical Performance Factors www ets lindgren com 905 STUB PROBE...
Страница 25: ...www ets lindgren com Typical Performance Factors 25 Preamplifier Gain 0 01 MHZ 3 GHZ...
Страница 26: ...26 Typical Performance Factors www ets lindgren com This page intentionally left blank...
Страница 49: ...www ets lindgren com EC Declaration of Conformity 49 Appendix A EC Declaration of Conformity...