Installation Procedure 

Esco Miri® Multi-Room Incubator 

 

 

 

ESCO MEDICAL 

MRI-6A10/XQ-1

 

 

 

Version 1.1

Содержание Miri MRI-6A10/XQ-1

Страница 1: ...Installation Procedure Esco Miri Multi Room Incubator ESCO MEDICAL MRI 6A10 XQ 1 Version 1 1 ...

Страница 2: ...CAL products should not hesitate to contact us if there are any unclear points or ambiguities in this manual CAUTION If the equipment is used in a manner not specified in this manual the safety of the user may be at risk and the equipment may be damaged Always use the equipment as outlined in this instruction manual COPYRIGHT This manual contains information that is subject to copyright All rights...

Страница 3: ...he Product 6 5 Manuals Supplied 7 5 1 The User Manual 7 5 2 The Validation Manual 7 5 3 The Maintenance Manual 7 6 Installation 8 6 1 Responsibilities 8 6 2 Before Installation 8 6 3 Preparing Installation 9 6 4 Bring To the Installation Site 9 6 5 Installation Procedure at the Site 10 6 6 User Training 10 6 7 After the Installation 11 7 Technical Assistance 12 ...

Страница 4: ... with the acquisition of your MRI 6A10 XQ 1 incubator We at ESCO MEDICAL are proud that you have chosen our product to serve your needs We hope to hear from you with feedback on this product that can benefit future product developments We will strive to serve you well Thank you ...

Страница 5: ...llation procedures is detailed in Chapter 5 and forwards Chapters from 2 5 can be skipped if prior knowledge about the product exist Chapters 5 and forward must be read chronologically 3 Explanation of Symbols Used NOTE Used to direct attention to a specific item DANGER Used when caution is needed ...

Страница 6: ... culture chambers and the lids The Incubator needs 100 CO2 and 100 N2 in order to be able to control the CO2 and O2 concentrations in the culture chambers The incubator has been primarily developed and optimized for tissue culturing with an overlay of either Paraffin or Mineral oil If open culture is used the user must switch the device to open culture mode Open culture may lead to evaporation and...

Страница 7: ... Manual Is intended to provide the trained technical personnel that is performing the installation the procedures and specifications for making sure the device is safe so that it can be taken into clinical use The manual also includes a guide for the validation tests that should be performed regularly 5 3 The Maintenance Manual Is intended to provide the schedules and the methods for the trained t...

Страница 8: ... have a clear understanding of the function performance testing and maintenance of the instrument Installation personnel must be updated as regards to alterations or additions to this document and the Installation test form that can be found in the validation manual in section 8 13 6 2 Before Installation 2 3 weeks before the Installation time is up the user owner at the Clinic is contacted via e ...

Страница 9: ...e sure it is the latest and current version only Fill out the following blank boxes in the form MRI 6A10 XQ 1 S N and Customer A service tool kit is checked for content before every installation trip to assure it contains the needed tools Always bring the latest versions of Firmware and the Data logging SW Bring these files on a memory stick with label to the service site 6 4 Bring to the Installa...

Страница 10: ...f the MRI 6A10 XQ 1 If there are any suggestions or complaints these must also be recorded in the installation test form 6 6 User Training Mains switch ON OFF Explain basic function of MRI 6A10 XQ 1 Incubation with multi room facility for storing the samples Explain temperature control in MRI 6A10 XQ 1 Direct heat transfer with heat ed lids Gas regulation on off Set point for temperature CO2 and O...

Страница 11: ...he date of Installation is written in the Instrument Overview file The date of Installation is also written in the Service schedule If the MRI 6A10 XQ 1 user or owner make inquiries about a written Installation report the completed and signed Installation test form must be sent to the clinic Any deviations complains suggestions from the Installation visit are reported in the CAPA system If a criti...

Страница 12: ...the company within ESCO MEDICAL which supplies them A copy of these terms and conditions is available only on request Contact your local ESCO MEDICAL representative for the most current information 2012 ESCO MEDICAL All rights reserved ESCO MEDICAL Esco Micro Pte Ltd 21 Changi South Street 1 Singapore 486777 Tel 65 6542 0833 Fax 65 6542 5732 6546 2913 Mail csis medical escoglobal com Web www escog...