MIRI

®

 II-12 Multiroom IVF incubators User Manual Rev. 4.0 

42 

 

 

Figure 18.3

 4-well dish with 3 buffers and media

 

 

 For calibration, at least two buffers are needed. We recommended using 3 buffers. 

One of the buffers should have a 7 pH. Any pH buffer can be used as the user's buffer 
levels can be set in the calibration dialogue window. If only one or two buffers are 
available, the system can still be used but with reduced accuracy.

 

 

The technique requires the user to be quick, as the pH starts to shift very quickly once the 
lid is opened. The optimal time to complete the procedure is tested to be 15 seconds, giving 
the same results as the continuous measurement described below. 

 

 

Press the “Calibrate” key:

 

 

 

Figure 18.4

 pH calibration screen view

 

 

Set the buffer levels with the (+) and (-) keys to correspond to the buffers used.

 

 

Before measuring in the culture media, calibrate the probe in 2 or 3 buffers. It is necessary 
to rinse the probe between each insertion.

 

 

Содержание MIRI II-12

Страница 1: ...USER MANUAL Rev 4 0 Date of revised 06 10 2021 Rx only MIRI II 12 Multiroom IVF Incubators For research use only...

Страница 2: ...uct is copyrighted and all rights are reserved by Esco Esco reserves the right to make periodic minor design changes without obligation to notify any person or entity of such change Sentinel is a regi...

Страница 3: ...rtial refund credit the product must not have been damaged It must be returned complete meaning all manuals cables accessories etc within 30 days of original purchase in as new and resalable condition...

Страница 4: ...ABUSE IS NOT COVERED UNDER THE WARRANTY Items such as cables and non serialized modules are not covered under this warranty This warranty gives you specific legal rights and you may have other rights...

Страница 5: ...ill be required to provide us with the serial number for your instrument as well as a valid reason for breaking the Quality Seal You should break this seal only after you have received factory authori...

Страница 6: ...4 7 Safety symbols and labels 14 8 Important safety instructions and warnings 16 8 1 Before installation 16 8 2 During installation 16 8 3 Post installation 17 9 Getting started 17 10 Mains connection...

Страница 7: ...es and measuring temperature 37 16 Pressure 39 16 1 CO2 gas pressure 39 16 2 N2 gas pressure 39 17 Firmware 40 18 pH measuring 40 19 SAFE Sense function 43 20 Data logging 43 20 1 Data logging tempera...

Страница 8: ...4 Gas concentration levels and consumption 60 33 1 5 Cosmetic 61 34 Validation on site 61 34 1 Mandatory equipment 61 34 2 Recommended additional equipment 62 35 Testing 62 35 1 Gas supply CO2 62 35...

Страница 9: ...37 3 In line HEPA filter for N2 gas 75 37 4 O2 sensor 75 37 5 CO2 sensor 76 37 6 UV light 77 37 7 Cooling fan 77 37 8 Internal gas pump 78 37 9 Proportional valves 78 37 10 Gas lines 78 37 11 Flow se...

Страница 10: ...tions given in this documentation may damage the unit injure operating personnel and or poor equipment performance Any internal adjustment modification or maintenance to this equipment must be un dert...

Страница 11: ...op and the bottom of each compartment is separated with a PET layer so that the lid temperature would not affect the bottom For validation purposes each compartment has a PT 1000 sensor built in The c...

Страница 12: ...y performing this the manufacturer can provide a valuable service to the customers MIRI II 12 A pH sensor port is part of the DAQ package The user can plug any standard BNC pH probe to the unit and me...

Страница 13: ...t the accompanying documents for important safety related information such as warnings and precautions that cannot be presented on the device itself for various reasons Do not use if the packing mater...

Страница 14: ...ncubator to guide the user User labels are shown below Table 7 1 Packing box and electrical safety labels Description Image Packing box label 1 If stored longer than the shelf life the unit must be re...

Страница 15: ...for use 7 Observe WEEE 8 Upper limit of temperature 9 Keep away from direct sunlight 10 Keep dry 11 Logo and serial number 12 Year of manufacture 13 Rx only 1 For research use only Table 7 3 Labels on...

Страница 16: ...user manual thoroughly before use 3 Always keep these instructions easily accessible near the device 8 2 During installation 1 Never place this unit on top of other equipment that gives off heat 2 Pla...

Страница 17: ...n or moisture or does not operate normally The MIRI II 12 multiroom IVF in cubators contain high voltage components that may be hazardous 3 Unplug this device during lightning storms or when unused fo...

Страница 18: ...the manufacturer s recommended guidelines 10 Mains connection The MIRI II 12 multiroom IVF incubators come with a detachable mains power cord The power cord is prepared for the country in which the un...

Страница 19: ...10 0 by infusing N2 gas Gas pressure for both inlets should be between 0 4 0 6 bar 5 80 8 70 PSI and it must be kept stable Always use a high quality pressure regulator that can be set with the requir...

Страница 20: ...reach the blastocyst stage VOC levels in the 0 5 ppm range will typically allow an acceptable blastocyst development and reasonable pregnancy rates but will likely result in a high percentage of misca...

Страница 21: ...d incorrectly will cause the unit to work not as intended This is dangerous The filter is removed by gently pulling it straight out using both hands Never run the MIRI II 12 multiroom IVF incubator if...

Страница 22: ...come back on automatically after 5 min It can be muted again Display panel Shows the information about the current status of the unit The display consists of a 7 touchscreen display 13 1 Activating t...

Страница 23: ...range from 25 0 C to 40 0 C The default temperature setpoint is 37 0 C To change the temperature setpoint please follow these instructions 1 Press one of the compartments boxes to change the temperatu...

Страница 24: ...rature setpoint is set then the value will be applied to all compartments If the MULTI temperature setpoint is set it means that the compartments temperature values are individual for each compartment...

Страница 25: ...ure below the setpoint for CO2 is chosen Figure 13 6 CO2 setpoint view 13 4 Changing the O2 concentration setpoint The O2 setpoint can be adjusted in the range from 5 0 to 10 0 The default O2 setpoint...

Страница 26: ...igure 13 7 Main display view 2 Press or keys to adjust the O2 setpoint In the picture below the setpoint for O2 is chosen Figure 13 8 O2 setpoint view 3 Press the SETUP key in the O2 setpoint page to...

Страница 27: ...nter the mode setting The key is located on the left bottom side of the main display Figure 13 11 Running mode key location on the main display Press the LOGGER key to enter the data logger The key is...

Страница 28: ...the top of the main display Figure 13 15 Report export key location on the main display 13 5 1 General menu page Press the SETTINGS key to enter the general menu page The key is located on the right...

Страница 29: ...s shown below Figure 13 19 Calibration of the T1 zone temperature view In the above picture zone T1 is chosen Press or keys to calibrate T1 The same applies to other temperature zones Press the key at...

Страница 30: ...overlay Place the dish in one of the designated spots indicated on the heating insert Proceed to validate if the lid temperature is precisely 0 2 C higher than the bottom temperature Stick a suitable...

Страница 31: ...CO2 concentration was measured to be 6 4 By pressing the button three times it will take time to change the display s CO2 concentration value but the offset value will change immediately in this case...

Страница 32: ...g mode page The running mode page is shown below Figure 13 22 Operation mode page view 1 Choose between the SINGLE or MULTI temperature setpoint If the SINGLE temperature setpoint is set then the valu...

Страница 33: ...ich indicates the alarm condition The audio pattern is 3 short beeps followed by a 3 second pause All alarms have the same audio pattern 14 1 Temperature alarms All 12 compartments can trigger the tem...

Страница 34: ...ture below the CO2 concentration is too low compared to the setpoint Figure 14 3 Low CO2 concentration alarm view on the main display The percentage of CO2 is too low CO2 concentration will appear in...

Страница 35: ...by blinking red when the alarm is muted 14 3 Gas pressure alarms 14 3 1 CO2 pressure alarm If the CO2 gas supply is not attached correctly or incorrect CO2 gas pressure is applied to the system the MI...

Страница 36: ...es above 0 7 bar 10 20 PSI it will trigger the alarm Figure 14 6 N2 gas pressure alarm view on the main display An audible alarm is also activated but it can be muted by pressing the alarm key If the...

Страница 37: ...he alarm is muted 14 6 Loss of power alarm If the incubator s power is disconnected an audio alarm will go on for approximately 4 seconds and the LED in the mute alarm key will flash Figure 14 9 The a...

Страница 38: ...tment 1 T1 T7 Compartment 2 T2 T8 Compartment 3 T3 T9 Compartment 4 T4 T10 Compartment 5 T5 T11 Compartment 6 T6 T12 Compartment 7 T13 T19 Compartment 8 T14 T20 Compartment 9 T15 T21 Compartment 10 T1...

Страница 39: ...gas pressure The CO2 pressure can be seen in the CO2 box on the main page as shown below Figure 16 1 CO2 pressure on the main display The CO2 pressure value is shown in bar The external pressure must...

Страница 40: ...ng to the maintenance plan 17 Firmware The firmware installed on your MIRI II 12 multiroom IVF incubator is upgradeable Whenever an important update is available it will be provided to our distributor...

Страница 41: ...esponding to a measurement done with an external device Otherwise the measurement will be incorrect as pH is a temperature dependent measurement All readings from the pH system and calibration dialogu...

Страница 42: ...system can still be used but with reduced accuracy The technique requires the user to be quick as the pH starts to shift very quickly once the lid is opened The optimal time to complete the procedure...

Страница 43: ...probe When choosing an electrode probe it is necessary to consider the probe s size as measurements will be made on either a 4 well dish or a droplet 19 SAFE Sense function There is a possibility to...

Страница 44: ...g a finger over it Zooming can be re peated in steps To get back to the original size press the Reset button 20 2 Data logging CO2 view By pressing the CO2 button the view will change to the CO2 graph...

Страница 45: ...ize Figure 20 4 Temperature concentration and pressure alarm view The Lid section has 12 rows in the MIRI II 12 multiroom IVF incubator Each row indicates a single lid opening case in a particular com...

Страница 46: ...procedure is recommended for routine processing and maintenance The combination of periodic cleaning procedures and disinfection procedures is recommended for event related concerns such as media spil...

Страница 47: ...time spray sterile water on the surfaces and wipe them with a sterile wipe 6 Inspect the device if it is visually clean consider it ready for use If the device is visually not clean go to step 3 and...

Страница 48: ...d designed for incubation of gametes and embryos with an overlay of either Paraffin or mineral oil The MIRI II 12 multiroom IVF incubator must not be irrigated Humidification of the MIRI II 12 multiro...

Страница 49: ...m one of the 12 gas sample ports on the unit s side using a suitable gas analyzer Figure 25 1 Gas sample ports Each sample port is directly connected to the corresponding compartment with the same num...

Страница 50: ...gases or if the power supply to the unit is suddenly cut the switch is indicating that the unit needs to be inspected by the user The connector can be connected either to a voltage source OR to a curr...

Страница 51: ...ed for easy reference during the incubation process The text can be wiped off with a cloth afterward Use only a suitable non toxic pen that allows the text to be erased later and will not damage the i...

Страница 52: ...29 Emergency procedures Total loss of power to or inside the unit Remove all the samples and place them in an alternative or backup device that is not affected by the problem Without the power source...

Страница 53: ...porary and the CO2 level is low keep the lids shut If the state is temporary and the CO2 level is high open a few lids to vent out some CO2 If the O2 level alarm goes off Usually no emergency procedur...

Страница 54: ...e power mains The system is switched off Switch the system on CO2 gas regulator is off Activate CO2 gas regulator by setting CO2 to ON in the menu No CO2 gas or wrong gas attached to CO2 gas input Che...

Страница 55: ...ds N2 gas pressure indicated red on the display No wrong N2 gas pressure in the system Check the N2 gas supply ensure that the pressure is stable at 0 6 bar 8 70 PSI If O2 regulation is not needed set...

Страница 56: ...consumption 500W Temperature range 25 0 40 0 C Gas consumption CO2 1 2 liters per hour Gas consumption N2 2 12 liters per hour CO2 range 3 0 10 0 O2 range 5 0 10 0 CO2 gas pressure input 0 4 0 6 bar...

Страница 57: ...Voltage fluctuations flicker emissions IEC 61000 3 3 Class A Table 32 2 Electromagnetic immunity Guidance and manufacturer s declaration electromagnetic immunity The MIRI II 12 multiroom IVF incubator...

Страница 58: ...rding to the equation applicable to the transmitter s frequency Recommended separation distance d 0 35 P d 0 35 P 80MHz to 800MHz d 0 7 P 800MHz to 2 5GHz Where P is the maximum the power output ratin...

Страница 59: ...stance applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflections from structures objects and people Medical Devices may be a...

Страница 60: ...sensor N2 concentration variation from setpoint within 0 2 absolute Pass IV Gas flow CO2 less than 2 l h Pass V Gas flow N2 less than 12 l h 33 1 2 Electrical safety An electric safety test is also ca...

Страница 61: ...e most comprehensive tests before the device is shipped to the customer there is no way to be sure that everything is still OK at the location when the device is set up Therefore in keeping with estab...

Страница 62: ...of 100 CO2 at 0 4 0 6 bar 5 80 8 70 PSI of pressure Measure the CO2 concentration in the gas supply by routing the gas line into a bottle without a lid and a suitably large opening Set the pressure f...

Страница 63: ...v max Oxides of Nitrogen NO NO2 2 5 ppm v v max each The non volatile residue particulates 10 ppm w w max The non volatile organic residue oil and grease 5 ppm w w max Phosphine 0 3 ppm v v max Total...

Страница 64: ...N2 gas can act as a simple asphyxiant by displacing air The User should make sure the N2 used is safe and moisture free Below is a list of some standard component concentrations Please note that the v...

Страница 65: ...ing device PASS The value must be 0 4 0 6 bar Please refer to the User manual sections for more information 35 5 Voltage supply The voltage on site must be verified Measure the output plug on the UPS...

Страница 66: ...nalyzer must have a gas return port connected to the multiroom IVF incubator i e another compartment Only measure while the value on the gas analyzer stabilizes Please refer to the 13 5 5 O2 calibrati...

Страница 67: ...on PASS all temperatures measured on the bottom of the compartments where the dishes are located must not deviate more than 0 1 C from the setpoint 35 9 Temperature check Compartment lids The second p...

Страница 68: ...st should be conducted with O2 regulation switched on and with N2 gas supply If the N2 is not available the test can be done without it Make sure that the Esco Medical data logger software is running...

Страница 69: ...the incubator Let the incubator run empty for at least 20 minutes before inserting a sample 35 12 Test documentation form The Installation report form must be completed with the tests passed status f...

Страница 70: ...ature check X CO2 gas concentration check X O2 gas concentration check X Check log for anomalies X CO2 gas pressure check X N2 gas pressure check X pH check X 36 1 Temperature check The temperature ch...

Страница 71: ...ion It is essential to have a high precision gas analyzer for CO2 and O2 available to do the test Please follow these simple rules while testing gas concentration Check the O2 gas setpoint Check the a...

Страница 72: ...re of 0 4 0 6 bar on the input N2 gas line This gas pressure must be held stable at any time For safety reasons this unit has a built in digital gas pressure sensor control that monitors the incoming...

Страница 73: ...ons given in the 33 The Validation guide section If problems are encountered contact Esco Medical or your local representative However to sustain the high performance level and avoid system errors the...

Страница 74: ...acement In addition to the active carbon component this capsule also has an integrated HEPA filter inside enabling it to remove particles and volatile organic compounds from the air being re circulate...

Страница 75: ...es found in the incoming N2 gas Failure to use the HEPA filter may cause damage to the high precision flow sensor calculate the amount of N2 gas entering the system and disturb the N2 regulator system...

Страница 76: ...sensor within 1 year from the date of the previous sensor installation Failure to change the oxygen sensor on time will result in low no regulation of O2 concentration Warranty void if wrong non origi...

Страница 77: ...ulb on time can result in contamination build up Warranty void if wrong non original UV light bulb is used Please refer to the service manual for replacement instructions 37 7 Cooling fan The cooling...

Страница 78: ...s or breakdowns Warranty void if wrong non original pump is used Please refer to the service manual for replacement instructions 37 9 Proportional valves The internal valves make gas regulation possib...

Страница 79: ...This sensor s lifetime is more than 3 years but Esco Medical recommends the sensor to be replaced once every 2 years for safety reasons Please follow these safety precautions when changing sensors Alw...

Страница 80: ...Always use the original filter contact Esco Medical or your local distributor for more details or ordering Change the filter once every year Failure to change the filter on time will result in low no...

Страница 81: ...als who will perform installation repair and or maintenance of the instrument must be trained by Esco Medical or at a qualified training center Experienced service technicians or embryologists conduct...

Страница 82: ...atest versions of firmware and data logging software Bring these files on a labeled memory stick to the service site 38 4 Bring the following to the installation site Installation report form Service...

Страница 83: ...nd calibration of temperature 13 External measurement and calibration of gas concentration 14 How to add and remove a sample 15 Demonstrate how to replace the VOC HEPA filter can be found in the 12 1...

Страница 84: ...n about this will be reported directly to QC or QA If the MIRI II 12 multiroom IVF incubator fails any of the Installation report form acceptance criteria or it in any way suffer from a severe error a...

Отзывы: