eschmann TD830 Скачать руководство пользователя страница 9

E-IM54f

P9/37

TD830

ELECTROSURGICAL UNIT

3.0  SAFETY NOTES & ALARMS

Attention to the following points is necessary
in order to reduce the risk of accidental burns
during the use of this electrosurgical unit. Other
safety notes and warnings are also given within
the text of this manual and these should be
noted before using this electrosurgical unit.

DO:

Use only Eschmann accessories, in particular active
and plate electrodes and cables which should
preferably be no more than 3 metres in length. All
Eschmann accessories are rated above the
corresponding maximum peak output voltage of this
electrosurgical unit (see graphs in section 5). For
reference Eschmann accessories can be used safely
at the following h.f. peak voltages:

Bipolar active - 400 V peak
Monopolar active - 4000 V peak
Monopolar plate electrode - 2250 V peak

Use bipolar techniques in preference to monopolar,
whenever possible.  For surgical procedures on parts
of the body having a relatively small cross-sectional
area, the use of bipolar techniques may be desirable
to avoid unwanted coagulation.

Seek approved qualified advice (e.g. cardiology
department) before using this electrosurgical unit on
patients with implanted pacemakers or other active
implants to avoid interference or damage to the implant.
Monitor such patients carefully.

Set power output controls to the minimum setting
before use and select the minimum power setting to
achieve the desired effect.

Check all cables and accessories routinely before
use. In particular, electrode cables and endoscopically
used accessories for possible damage to the insulation.

Ensure the entire area of the plate electrode is reliably
attached to the patient’s body and as close to the
operating field as possible.

Store temporarily unused active electrodes such that
they are isolated from the patient [e.g. use a quiver
(REF 83-186-38) to hold active accessories when not
in use].

DO NOT:

Do not use uninsulated forceps, monopolar or bipolar.

Do not place monitoring electrodes close to the
operating site.  When high frequency surgical
equipment (i.e. electrosurgical equip-ment) and
physiological monitoring equip-ment are used
simultaneously on the same patient, any monitoring
electrodes should be placed as far as possible from
the surgical electrodes.  Needle monitoring electrodes
are not recommended.  In all cases, monitoring
systems incorporating high frequency current limiting
devices are recommended.

Do not allow active cables to drape across or contact
the patient's body or contact the cables or leads of
other equipment.

Do not use flammable anaesthetics, flammable
solvents or oxidizing gases such as nitrous oxide
(N

2

O) or oxygen if the surgical procedure is carried

out in the region of the thorax or head, unless these
agents are sucked away.

Do not allow the patient's body to touch conductive
objects.  The patient should not come into contact with
metal parts which are earthed or which have an
appreciable capacitance to earth, e.g. operation table,
supports, etc.  The use of antistatic sheeting is
recommended for this purpose.

Do not use hook type active cables (with or without
adapters designed for use with hook type cables) use
4 mm or 8 mm plug type active cables.

Do not reuse disposable plate electrodes.

Do not rely solely on surgical gloves to provide
insulation.

Do not allow ‘skin-to-skin’ contact (e.g. between the
arms and body of the patient), this can be avoided, for
example by the insertion of adequate dry gauze.

Содержание TD830

Страница 1: ...TD830 ELECTROSURGICAL UNIT Instructions for use E IM54f 698257...

Страница 2: ...Sussex BN15 8TJ England Tel 44 0 1903 765040 Fax 44 0 1903 875711 Overseas Customers Contact your local distributor In case of doubt contact Eschmann Equipment Patents and Trade marks The ESCHMANN nam...

Страница 3: ...ns 14 Electrosurgical interference 14 Using two electrosurgical units 15 Using a flexible endoscope 15 Servicing 15 References 15 5 OPERATION Routine check 16 Connection of accessories 16 Monopolar 16...

Страница 4: ...is most important that fuses of the correct type size and rating are installed see Technical Data CAUTION Read these Instructions for use carefully before using this electrosurgical unit and note ALL...

Страница 5: ...25V min T10A Fuse type 20 mm Bipolar Carrier frequency 785kHz nominal square wave Power control Variable pulse group modu lation set by front panel control Amplitude set by micro macro range buttons L...

Страница 6: ...rd edition General classification Class 1 Type CF Defibrillator proof Drip proof IPX 1 Patient leakage risk current always less than 100 microamps to earth ground from all patient circuits at 230V 50H...

Страница 7: ...he symbol on the rear panel serial plate warns the user to read the accompanying documents the Instructions for use The symbol above sockets 25 26 27 and 28 denotes connection socket for a two button...

Страница 8: ...Plate voltage monitor alarm has been activated EPM Indicates the Excess power monitor alarm has been activated Indicates the Multiple activation alarm has been activated Indicates an internal fault ha...

Страница 9: ...heplateelectrodeisreliably attached to the patient s body and as close to the operating field as possible Store temporarily unused active electrodes such that they are isolated from the patient e g us...

Страница 10: ...f the delivered power exceeds the allowed value by more than a certain margin Patient Earth Monitor PEM 3 5 The PEM circuit detects a low impedance path between the patient and earth It also reduces t...

Страница 11: ...f the bipolar forceps and the tissue held between them is directly in view of the surgeon at all times 4 7 A plate electrode is not required and safety is increased even further as two power range out...

Страница 12: ...ined throughout the surgical procedure The following points should also be noted Do not apply any plate electrode over bony protuberances metal implants folds of skin scar tissue hairy areas any form...

Страница 13: ...ited number of re sterilization cycles Regular testing and inspection is therefore essential Do not sterilize and reuse single use types 4 32 Do not connect footswitches to the USER 1 output of the el...

Страница 14: ...te electrode are kept a few inches away from the pacemaker pulse generator and its electrode leads while to avoid burns to the myocardium at the electrode site or possible ventricular fibrillation the...

Страница 15: ...the following notes should be considered when writing a procedure for the use of this electrosurgical unit with an endoscope a To avoid the risk of potential burns caused by h f voltages on the body o...

Страница 16: ...a supply of new plate electrodes is readily available Plate electrode With reusable types examine the platecableandplugforsignsofwear Ensurethatthe insulationoftheconnectingcableisingoodcondition Acce...

Страница 17: ...olar electric socket 37 or a white pneumatic footswitch REF 83 581 68 Fig 10 to the bipolar pneumatic socket 36 Do not connect both types of footswitch at the same time 5 9 If the electrosurgical unit...

Страница 18: ...uld be checked regularly in accordance with the Service Manual see section 1 8 OPERATING INSTRUCTIONS Note For numbers in brackets refer to Illustration 11 which folds out at the end of this manual Un...

Страница 19: ...obtained with the normal power setting check the following before increasing power The plate electrode is correctly positioned and is in contact with the patient Theplateelectrodecableisintact andispl...

Страница 20: ...non contact coagulation of tissue that tends to adhere to active electrodes Note Spray coag indicated with the bright green LED is not recommended for laparoscopic and endoscopic procedures where its...

Страница 21: ...de or press macro bipolar selector touch button 5 to select macro bipolar mode The green indicators 2 or 4 will illuminate to show which bipolar mode has been selected 5 49 Set power output control 1...

Страница 22: ...nit is in standby mode only the bipolar display will flash 5 57 If more than one output demand is received at the same time such as two footswitches or a footswitch and a fingerswitch button being pre...

Страница 23: ...s been activated Remedy i If unit PVM alarm activates check for accidental direct contact between the active electrode and an earthed object ii Check that the plate electrode is correctly attached to...

Страница 24: ...ation 40 to 60 TCRE TCAE Blend Normal endometrium 55 to 70 Transcervical Resection of Specialist cut Spongio endometrium 40 to 60 the Endometrium Ablation Blend Suspect fibroid inv 65 to 80 Spray coag...

Страница 25: ...Power output watts Load resistance 200 ohms MONOPOLAR CUT POWER v CONTROL 0 50 100 150 200 250 300 0 25 50 75 100 125 150 175 200 225 250 Output control setting typical pow er Power output watts Load...

Страница 26: ...0 ohms MONOPOLAR BLEND POWER v CONTROL 0 40 80 120 160 200 0 20 40 60 80 100 120 140 160 180 200 Output control setting typical power Power output watts Load resistance 400 ohms MONOPOLAR BLEND DIAGRA...

Страница 27: ...ance 200 ohms MONOPOLAR SPECIALIST CUT POWER v CONTROL 0 50 100 150 200 250 300 350 0 30 60 90 120 150 180 210 240 270 300 Output control setting typical pow er Power output watts Load resistance 200...

Страница 28: ...MONOPOLAR PINPOINT COAG POWER v CONTROL 0 20 40 60 80 100 120 0 10 20 30 40 50 60 70 80 90 100 Output control setting typical pow er Power output watts Load resistance 400 ohms MONOPOLAR PINPOINT COA...

Страница 29: ...alue up to 10 Peak output voltage volts Open circuit MONOPOLAR SPRAY COAG VOLTAGE v CONTROL 0 500 1000 1500 2000 2500 3000 3500 4000 0 7 5 15 22 5 30 37 5 45 52 5 60 67 5 75 Output control setting typ...

Страница 30: ...tput voltage volts Open circuit BIPOLAR MACRO POWER v CONTROL 0 10 20 30 40 50 0 1 2 3 4 5 6 7 8 9 10 Output control setting numerical value up to 10 Power output watts Load resistance 100 ohms BIPOLA...

Страница 31: ...pow er Peak output voltage volts Open circuit BIPOLAR MICRO POWER v CONTROL 0 4 8 12 16 0 1 5 3 4 5 6 7 5 9 10 5 12 13 5 15 Output control setting typical pow er Power output watts Load resistance 10...

Страница 32: ...Eschmann trained personnel otherwise the warranty for the equipment could be infringed 6 5 If the unit is to be stored for any length of time it should be plugged into the mains and switched on to ch...

Страница 33: ...sposable 8 83 588 85 4 2mm pack of 5 8 REF Description Illustration No Colposcopy electrodes LLETZ cone biopsy 83 583 51 large loop 12mm pack of 5 8 83 584 19 large loop 15mm pack of 5 8 83 584 27 lar...

Страница 34: ...34 37 E IM54f 83 122 42 83 122 46 83 122 49 83 207 56 83 122 93 83 109 39 83 109 47 83 581 09 83 581 13 83 581 17 83 581 25 83 140 20 4mm 8mm 83 141 01 83 141 36 83 141 28 83 141 44 1 2 3 4 5 83 109 7...

Страница 35: ...8 m m 6 7 8 83 582 14 83 582 27 83 582 30 83 582 49 83 588 69 83 585 04 Disposable 83 585 11 Disposable 83 582 22 83 583 11 83 585 23 Disposable 83 588 77 83 588 80 83 583 51 83 584 19 83 584 27 83 5...

Страница 36: ...P36 37 E IM54f 9 1 0 83 700 01 83 703 03 83 705 08 83 707 02 83 701 09 83 704 00 83 706 05 83 708 18 83 702 06 83 709 07 83 581 76 83 581 68...

Страница 37: ......

Страница 38: ...Eschmann Equipment Peter Road Lancing West Sussex BN15 8TJ England Tel 44 0 1903 753322 Fax 44 0 1903 875789 www eschmann co uk...

Страница 39: ...for use with active electrode socket 26 black 26 4mm active electrode socket USER 1 footswitch or fingerswitch activation red 27 Fingerswitch sockets for use with active electrode socket 28 black 28...

Отзывы: