Tips
1. You must stay still until you hear the beep for measurement completion. The thermometer
cannot be moved away from the position of measuring temperature before the measurement
is completed. Doing so may cause inaccuracies.
2.
When taking the temperature, the product should be pointed to the center of the forehead,
above the brow center, and kept vertical. The measuring area should not be covered by hair.
The distance between the product and the forehead is recommended to be about 3 – 5 cm.
Note
:
If the forehead cannot be used for measurement, use a body surface such as the chest, or
abdomen.
3. When a person comes from a place where there is a significant difference of environment
temperature, the person being measured should stay in the new environment (where they are
being measured) for at least 5 minutes; otherwise the measured result will be affected.
The measurement result will be on the low end if a fever patient’s forehead has sweat, or if
they have had other quantitative cooling measures; this should be avoided.
If the product is stored in a place where there are environmental temperature differences, it
should be placed in the new environment for 20 minutes prior to use.
When taking a measurement, the environment should be stable. Do not take temperature in
the outlet of a fan, air conditioner, or any large airflow.
Do not use the product in direct sunlight.
Settings
Switching Modes
Press the “measure switch/trigger” once. When the LCD is lit, press the “SET” button to
change between “Body” and “Surface” modes.
High Fever Alarm Temperature Setting
Press the “measure switch/trigger” once. When the LCD is lit, press and hold the “SET”
button until you see F-1, and then release the button. Press the “SET” button once until you
see F-2. You can now press the UP or DOWN arrow buttons to adjust the target temperature
for the fever alarm. Press the “SET” button until the LCD powers off to save your setting.
Default setting is 100.4°F (38°C).