Chapter 3
The magnetic unit and patient’s seat are located inside the controlled
access area.
Peripheral equipment
The peripheral equipment used inside the controlled access area must be
safe for use in MR environments.
This equipment guarantees that:
☛
no known risks are generated in any MR environment, or when used in
specific conditions
☛
the quality of diagnostic information supplied by the O-scan system is
not altered by peripheral equipment
☛
operation of peripheral equipment is not altered by that of the O-scan
system.
Peripheral equipment labeled as MR safe or MR conditional also
guarantees minimized probability of induced currents caused by
connection of the equipment and the RF transmission coil of the O-scan
system that can generate excessive heating or burns of the patient's skin.
As the patient's skin is conductive, closed conductive loops can form
between the patient and terminals (e.g. ECG electrodes) or cables of the
peripheral equipment, in which induced currents circulate with the risks
mentioned above.
W a r n i n g
Do not allow any persons, who have the conditions and
characteristics for which exposure to the O-scan magnetic and RF
fields is contraindicated, to enter the 0.5 mT restricted area.
W a r n i n g
Users are advised to affix the warning and prohibition labels provided
with the equipment on all of the doors of the O-scan examination
room.
Users are advised to mark the 0.5 mT line on the floor around the
O-scan installation area using the yellow adhesive tape that is
provided with the equipment.
Содержание O-scan eXP
Страница 1: ......
Страница 2: ......
Страница 3: ...O scan User Manual 350004900 Rev 07 Esaote S p A Genoa...
Страница 9: ...350004900 Rev 07 7 10...
Страница 10: ...8 10...
Страница 12: ...10 10...
Страница 20: ...8 8...
Страница 32: ...8 8 Chapter 2...
Страница 51: ...350004900 Rev 07 19 48 Instructions for Safe Use fig 3 1 Top view...
Страница 52: ...20 48 Chapter 3 fig 3 2 Side view...
Страница 80: ...48 48 Chapter 3...
Страница 102: ...22 22 Chapter 4...
Страница 105: ...350004900 Rev 07 3 20 Technical Description fig 5 1 Equipment components...
Страница 122: ...20 20 Chapter 5...
Страница 126: ...4 20 Chapter 6 fig 6 3 The O scan Premium and O scan eXP patient seat with armrests fig 6 4 Armrest detail...
Страница 142: ...20 20 Chapter 6...
Страница 155: ...350004900 Rev 07 7 8 Coils fig 8 5 Spherical uniform phantom fig 8 6 Cylindrical uniform phantoms...
Страница 156: ...8 8 Chapter 8 fig 8 7 Geometric phantom fig 8 8 Centering support for the spherical phantom diameter 140 mm...
Страница 160: ...4 10 Chapter 9 fig 9 4 Lumbar support cushion fig 9 5 Positioning of the lumbar support cushion on the seat...
Страница 161: ...350004900 Rev 07 5 10 Soft cushion Set fig 9 6 Use of lumbar support cushion...
Страница 184: ...10 10 Chapter 11...
Страница 232: ...4 4 Chapter 17...
Страница 236: ...4 4 Chapter 18...