P R E P A R I N G T H E S Y S T E M
MyLab
- G E T T I N G S T A R T E D
4 - 4
Fig. 4-3: C) Non-medical use room
A room not designed for medical treatment, for example, an office or a
storage room.
Possible configurations:
MyLab
+ auxiliary device complying to IEC 60601 in area A
- No additional safety requirements.
MyLab
+ auxiliary device not complying to IEC 60601
(complying to IEC XXX
1
) in area A
- Auxiliary device must be powered through a safety
insulation transformer complying to IEC 60601.
MyLab
+ auxiliary device not complying to IEC 60601
(complying to IEC XXX
) in area B or area C connected by
WiFi or Ethernet cable
- No additional safety requirements.
MyLab
+ auxiliary device not complying to IEC 60601 in area
B or area C connected by cable (USB, HDMI,...)
- Auxiliary device must be powered through a safety
insulation transformer complying to IEC 60601.
NOTE
Auxiliary Devices must be approved by Esaote. Auxiliary Devices must also
comply with EN 60601-1-2 safety standard and subsequent amendments or
the electromagnetic compatibility.
1.
IEC XXX stands for standards such as: IEC 60601 for medical devices, IEC 60950 for
information technology equipment etc.
Содержание MyLab 9VET
Страница 1: ...Rev 05 February 2020 MyLab 9VET GETTING STARTED 350041300 ...
Страница 2: ......
Страница 5: ...MyLab G E T T I N G S T A R T E D v GETTING STARTED EC Declaration of Conformity ...
Страница 6: ...MyLab G E T T I N G S T A R T E D vi RED Declaration of Conformity ...
Страница 7: ...MyLab G E T T I N G S T A R T E D vii GETTING STARTED ...
Страница 8: ...MyLab G E T T I N G S T A R T E D viii ...
Страница 14: ...I N D E X MyLab G E T T I N G S T A R T E D xiv ...
Страница 34: ...A D D I T I O N A L I N F O R M A T I O N O N S A F E T Y MyLab G E T T I N G S T A R T E D 2 8 ...
Страница 46: ...S Y S T E M O V E R V I E W MyLab G E T T I N G S T A R T E D 3 12 ...
Страница 116: ...T E C H N I C A L S P E C I F I C A T I O N S MyLab G E T T I N G S T A R T E D 9 12 ...