15
Troubleshooting
Fault:
The hoist does not react
when the hand control
buttons UP/DOWN are
used.
Possible cause:
Emergency stop is activated.
Solution:
Release the emergency stop (see section 2.3)
Possible cause:
Defect hand control.
Solution:
Try using another hand control. Alternative is to
get a new hand control from the dealer and
Winncare Nordic ApS.
Possible cause:
Flat battery.
Solution:
Charge battery (see section 2.4).
If this does not help replace batteries
Fault:
The lifting arm can
travel up or down not
both.
The electrical leg
spreading can only
move in or out not
both.
Possible cause:
Fault in the control box.
Solution:
Replace the control box.
Possible cause:
Defective hand control.
Solution:
Try using another hand control. Alternative is to
get a new hand control from the dealer and
Winncare Nordic ApS.
Fault:
The lifting actuator is
noisy or vibrates
abnormally
Possible cause:
Lifting actuator is defective.
Solution:
Replace the lifting actuator.
Possible cause:
Flat battery.
Solution:
Charge battery (see section 2.4).
Fault:
Does not charge
Possible cause:
Charger is not connected to the main supply.
Solution:
Connect charger.
Possible cause:
Solution:
Possible cause:
Emergency stop is activated.
Release the emergency stop (see section 2.3)
Defective control box.
Solution:
Try another control box. Order a new from
Winncare Nordic ApS or the dealer.
Possible cause:
Defective battery.
Solution:
Replace battery.