How to use
Clean and disinfect the product before using for the first time
and before every reuse.
The product can be sterilized. The user must decide if steriliza-
tion is necessary or not. National guidelines and local hygiene
regulations can be used to help make decisions.
WARNING! Do not kink the cable.
WARNING! Protect the product from all physical damage! Do not
throw! Do not apply force of any kind!
WARNING! Risk of injury! Do not activate the electrosurgical unit
while you are connecting the cable.
Connect the neutral electrode to the cable. Then plug the connec-
tor on the cable into a suitable socket on the electrosurgical unit.
4
Cleaning, disinfection, sterilization
4.1 Safety instructions for processing
May be cleaned in an ultrasonic bath.
Never use sharp objects for cleaning.
Disinfectants must be rinsed off thoroughly after use.
Maximum air pressure during drying: 2 bar.
Maximum water pressure during rinsing: 2 bar.
During machine cleaning and disinfection, do not exceed 95 °C.
Observe the specifications from the manufacturer of the deter-
gent and disinfectant and in the chapter “Overview of validated
procedures“.
During sterilization, do not exceed 138 °C.
Do not sterilize in hot air.
OBJ_DOKU-197684-002.fm Seite 21 Dienstag, 8. September 2020 9:57 09
Содержание 20194-070
Страница 28: ...28 OBJ_DOKU 197684 002 fm Seite 28 Dienstag 8 September 2020 9 57 09 ...
Страница 124: ...124 OBJ_DOKU 197684 002 fm Seite 124 Dienstag 8 September 2020 9 57 09 ...
Страница 136: ...136 OBJ_DOKU 197684 002 fm Seite 136 Dienstag 8 September 2020 9 57 09 ...
Страница 160: ...160 OBJ_DOKU 197684 002 fm Seite 160 Dienstag 8 September 2020 9 57 09 ...
Страница 172: ...172 OBJ_DOKU 197684 002 fm Seite 172 Dienstag 8 September 2020 9 57 09 ...
Страница 196: ...196 OBJ_DOKU 197684 002 fm Seite 196 Dienstag 8 September 2020 9 57 09 ...
Страница 218: ...218 OBJ_DOKU 197684 002 fm Seite 218 Dienstag 8 September 2020 9 57 09 ...
Страница 219: ...OBJ_DOKU 197684 002 fm Seite 219 Dienstag 8 September 2020 9 57 09 ...