103
16 Alarms and
Troubleshooting
The patch pump system has a
comprehensive safety system to check if
abnormal situations require immediate
attention. The system will send
notification alarms using sound, LEDs, or
vibrations as well as provide information
on the PDA display.
The insulin pump system contains two
parts ‒ the patch pump detects and sends
the alarms, and the PDA receives the
alarms and notifies the user.
The insulin pump system only contains
medium and low priority alarms ‒ no
high priority alarms (as defined by ISO
standards).
Medium priority
alarms occur when
the delivery function has stopped due
to technical failure, requiring the user to
intervene in the operation of the pump,
change the pump, or possibly inject insulin
manually to control BG levels.
Low priority alarms
alert the user
that something will occur in a short
period of time but insulin delivery
continues normally. Users should be
aware of this information and make plans
in advance to ensure that treatment
continues reliably.
There are no alarms to indicate old
insulin or to detect insulin leaks. The
user should be aware if these situations
happen.
Note
If the insulin pump and PDA start up
normally without alarms, the alarms
system is working properly.
Note
Note
Содержание MTM-I
Страница 1: ...Insulin Patch Pump System MTM I Insulin Patch Pump System User Manual TM...
Страница 2: ......
Страница 23: ...EquilTM Insulin Patch Pump System Figure 13 Figure 14 Figure 15 box Figure 15 that can be used to input the...
Страница 33: ...EquilTM Insulin Patch Pump System Time and Date Display Area Information Display Area Unlock Area Note Note...
Страница 45: ...EquilTM Insulin Patch Pump System Figure 45 1 Figure 46 2 47...
Страница 62: ...EquilTM Insulin Patch Pump System 57 Figure 70...
Страница 65: ...EquilTM Insulin Patch Pump System Figure 73 3 Figure 74 4 XXXXXX...
Страница 67: ...EquilTM Insulin Patch Pump System Figure 77 6 Note...
Страница 68: ...EquilTM Insulin Patch Pump System 63 Figure 78...
Страница 71: ...EquilTM Insulin Patch Pump System...
Страница 73: ...EquilTM Insulin Patch Pump System 4...
Страница 85: ...EquilTM Insulin Patch Pump System Figure 100 3 Note...
Страница 87: ...EquilTM Insulin Patch Pump System Figure 101 10 10 Settings Figure 102...
Страница 107: ...EquilTM Insulin Patch Pump System Figure 119 Figure 120 Note...
Страница 117: ...EquilTM Insulin Patch Pump System 17 6 X Ray MRI and CT Scans 17 7 Precautions 17 8 Wireless Connection...
Страница 119: ...EquilTM Insulin Patch Pump System 17 11 Storage 17 12 Other Considerations...
Страница 129: ...EquilTM Insulin Patch Pump System 19 Appendix 19 1 Symbols LOT SN IVD CTRL STERILE EO IPX4...