5-38
Issue 05, 02/00
Krautkramer USN 52R/USN 52L
Viewing stored readings
You can select each measurement location and have
the values displayed which are stored there. The follow-
ing appear as measurement location status:
EMPTY:
no measurement stored
x.xx mm:
the stored values in the active units of measure
ERROR:
erroneous measured value. Storing must be repeated.
Operating in the A-scan mode:
Measurement location number (LOC) and value (VAL)
are displayed in the measurement line.
– With
T
or
V
select the required measurement
location. The value stored there appears at
VAL.
Operation from the second operating level:
– Select the function group containing the Data Logger.
– Using
N
or
O
select the required measurement
location when in
LOCATION.
The value stored there
is displayed in
VALUE
.
Clearing the measured values
You can clear the displayed reading and, if necessary,
store a new value at this measurement location.
Operation in the A-scan mode:
– Select the value to be cleared as described above.
– Simultaneously press
U
and keep the keys
pressed for about 3 seconds.
When
VAL
: EMPTY appears, then the measured value
is deleted.
– If necessary, press
A
in order to store a new value
at this measurement location.
Operation from the second operating level:
– Select the value to be deleted, as described above.
– When in
VALUE,
simultaneously press
M
and keep
the keys pressed for about 3 seconds.
When
VALUE
: EMPTY appears, the measured value is
deleted.
– Press
A
in the A-scan mode to store a new value at
this measurement location.
Operation
Storing measured values - Data Logger
5-38
Issue 05, 02/00
Krautkramer USN 52R/USN 52L
Viewing stored readings
You can select each measurement location and have
the values displayed which are stored there. The follow-
ing appear as measurement location status:
EMPTY:
no measurement stored
x.xx mm:
the stored values in the active units of measure
ERROR:
erroneous measured value. Storing must be repeated.
Operating in the A-scan mode:
Measurement location number (LOC) and value (VAL)
are displayed in the measurement line.
– With
T
or
V
select the required measurement
location. The value stored there appears at
VAL.
Operation from the second operating level:
– Select the function group containing the Data Logger.
– Using
N
or
O
select the required measurement
location when in
LOCATION.
The value stored there
is displayed in
VALUE
.
Clearing the measured values
You can clear the displayed reading and, if necessary,
store a new value at this measurement location.
Operation in the A-scan mode:
– Select the value to be cleared as described above.
– Simultaneously press
U
and keep the keys
pressed for about 3 seconds.
When
VAL
: EMPTY appears, then the measured value
is deleted.
– If necessary, press
A
in order to store a new value
at this measurement location.
Operation from the second operating level:
– Select the value to be deleted, as described above.
– When in
VALUE,
simultaneously press
M
and keep
the keys pressed for about 3 seconds.
When
VALUE
: EMPTY appears, the measured value is
deleted.
– Press
A
in the A-scan mode to store a new value at
this measurement location.
Operation
Storing measured values - Data Logger
Содержание USN 52L
Страница 7: ...0 6 Issue 05 02 00 Krautkramer USN 52R USN 52L ...
Страница 8: ...Krautkramer USN 52R USN 52L Issue 05 02 00 1 1 Introduction 1 Introduction 1 ...
Страница 19: ...1 12 Issue 05 02 00 Krautkramer USN 52R USN 52L ...
Страница 25: ...2 6 Issue 05 02 00 Krautkramer USN 52R USN 52L ...
Страница 26: ...Krautkramer USN 52R USN 52L Issue 05 02 00 3 1 Setting into operation 3 Setting into operation 3 ...
Страница 32: ...Krautkramer USN 52R USN 52L Issue 05 02 00 4 1 Basics of operation 4 Basics of operation 4 ...
Страница 45: ...4 14 Issue 05 02 00 Krautkramer USN 52R USN 52L ...
Страница 46: ...Krautkramer USN 52R USN 52L Issue 05 02 00 5 1 Operation 5 Operation 5 ...
Страница 117: ...5 72 Issue 05 02 00 Krautkramer USN 52R USN 52L ...
Страница 118: ...Krautkramer USN 52R USN 52L Issue 05 02 00 6 1 Documentation 6 Documentation 6 ...
Страница 123: ...6 6 Issue 05 02 00 Krautkramer USN 52R USN 52L ...
Страница 124: ...Krautkramer USN 52R USN 52L Issue 05 02 00 7 1 Servicing and maintenance 7 Servicing and maintenance 7 ...
Страница 127: ...7 4 Issue 05 02 00 Krautkramer USN 52R USN 52L ...
Страница 128: ...Krautkramer USN 52R USN 52L Issue 05 02 00 8 1 Interfaces Peripherals 8 Interfaces Peripherals 8 ...
Страница 150: ...Krautkramer USN 52R USN 52L Issue 05 02 00 9 1 Specifications 9 Specifications 9 ...
Страница 155: ...9 6 Issue 05 02 00 Krautkramer USN 52R USN 52L ...
Страница 156: ...Krautkramer USN 52R USN 52L Issue 05 02 00 10 1 Appendix 10 Appendix 10 ...
Страница 165: ...10 10 Issue 05 02 00 Krautkramer USN 52R USN 52L ...
Страница 166: ...Krautkramer USN 52R USN 52L Issue 05 02 00 11 1 Changes 11 Changes 11 ...
Страница 168: ...Krautkramer USN 52R Issue 05 02 00 12 1 Index 12 Index 12 ...
Страница 176: ...Option DGS Operating Manual ...
Страница 179: ...DGS 4 Issue 02 03 99 Krautkramer USN 52R USN 52L Option DGS ...
Страница 180: ...Krautkramer USN 52R USN 52L Option DGS Issue 02 03 99 DGS 5 Introduction 1 Introduction 1 ...
Страница 184: ...Krautkramer USN 52R USN 52L Option DGS Issue 02 03 99 DGS 9 Overview of the DGS menu 2 Overview of the DGS menu 2 ...
Страница 191: ...DGS 16 Issue 02 03 99 Krautkramer USN 52R USN 52L Option DGS ...
Страница 192: ...Krautkramer USN 52R USN 52L Option DGS Issue 02 03 99 DGS 17 Operating the DGS 3 Operating the DGS 3 ...
Страница 203: ...DGS 28 Issue 02 03 99 Krautkramer USN 52R USN 52L Option DGS ...
Страница 204: ...Krautkramer USN 52R USN 52L Option DGS Issue 02 03 99 DGS 29 Codes for remote control 4 Codes for remote control 4 ...
Страница 207: ...DGS 32 Issue 02 03 99 Krautkramer USN 5 USN 52L Option DGS ...