Instruction Manual
PS 8000 DT Series
Date: 03-05-2016
Parameter
:
Broad ID
Default:
0x7FF
Settings:
0x000...0x7FF (0...2047)
Function: Adjusts the broadcast ID (BCID) for the CAN ID sys-
tem with three IDs (Vector compatible, dbc files). This extra ID
is a fourth ID for the device which can be used for broadcast
messages to multiple units on a bus. Purpose if this ID is to
adjust it to the same value on all units that are targeted to be
controlled simultaneously by set values or device conditions.
Display can be switched from decimal to hexadecimal by push-
ing any of the rotary knobs.
Only available, if
ID Sys
=
Vector
has been selected. See
below at parameter
ID Sys
.
Parameter
:
RID
Default:
0
Settings:
0...31
Function: Selects the relocatable identifier segment (RID). Refer
to CAN terminology or instruction manual of the IF-C1 CAN
interface card for further information.
Parameter
:
Bus term
Default:
yes
Settings:
yes, no
Function: activates/deactivates the bus termination resistor of
the CAN interface card. This is required if the device is at the
end of the bus.
Parameter
:
ID Sys
Default:
Vector
Settings:
Vector, normal
Function: Selects the CAN ID system. With „
Normal
“, the
former, old CAN ID system with two CAN IDs per unit is used,
where the IDs are built from „
Device node
“ (see above) and
„
RID
“ (see above) . Also see the external user manual for the in-
terface cards regarding the calculation scheme of the CAN IDs.
The other ID system, selected with „
Vector
“, uses three CAN
IDs per unit and it thus enables the use of so-called DBC files to
implement the device into Vector company software. By select-
ing this ID system, two ID related settings (see above) become
active, where the user adjusts a base ID that defines the three
CAN IDs, plus a broadcast ID (if used).
Following settings only with
RS232 interface IF-R1
:
Parameter
:
Baud
Default:
57600
Settings:
9600, 19200, 38400, 57600
Function: Selects the serial transmission baud rate (in baud).
Further parameters for the RS232 are not configurable, but
defined as follows:
Parity = odd
Stop bits = 1
Data bits = 8
and have to be set to the same configuration at the PC.
Following settings only with
Profibus interface IF-PB1
:
Parameter
:
Profibus
Default:
1
Settings:
1-125
Function: Defines the Profibus address of the device. This
address is used apart from the device node to implement and
access the unit on a field bus system.
Operating the device
8. Device setup
The device setup is intended to adjust certain operation param-
eters. It can only be accessed while the output is off. Push both
pushbuttons of the rotary knobs (also see section 6.3) simultane-
ously for longer than 2 s. Leaving the setup and storage of the
settings is done the same way. Three elementary settings are
always available, see below. Other settings are only available
if a digital interface card is equipped.
Elementary settings
:
Parameter
:
AutoPwrOn
Default:
on
Settings:
on, off
Function: „
on
“ --> device restores the last output state and set
values of last time the device was switched off or a blackout
occurred. The purpose is to make the unit automatically continue
operation after a blackout.
„
off
“ --> the output is switched on and the set values of U,
UVL and I are set to 0%, P to 100% and OVP to 110%, every
time the unit is powered.
Parameter
:
AI range
Default:
0-10
Settings:
0-5, 0-10
Function: selects the control voltage range to use with the
analogue interface. Further details in section 10.
Parameter
:
Contrast
Default:
70
Settings:
50...100
Adjusts the contrast of the LCD display.
For
all
interface cards this setting applies:
Parameter
:
Device node
Default:
1
Settings:
1...30
Function: Selects the device‘s address (i.e. device node, term
taken from the CAN terminology). When using the device on a
bus system (CAN or GPIB), every device must have a unique
address!
Following settings only with
CAN interface IF-C1:
Parameter
:
Baud
Default:
100 k
Settings:
10 k,
25 k,
50 k,
100 k,
125 k,
250 k,
500 k,
1M
Function: Selects the CAN transmission baud rate.
Parameter
:
Base ID
Default:
0x000
Settings:
0x000...0x7FC (0...2044)
Function: Defines the base ID (BAID) for the CAN ID system with
three IDs (Vector compatible, dbc files). Three IDs are reserved
for a device, based upon the adjusted base ID. Thus this values
is only adjustable in steps of four. Display can be switched from
decimal to hexadecimal by pushing any of the rotary knobs.
Only available, if
ID Sys
=
Vector
has been selected. See
below at parameter
ID Sys
.