![EPS Bio Technology EMV3.1 Скачать руководство пользователя страница 53](http://html1.mh-extra.com/html/eps-bio-technology/emv3-1/emv3-1_user-manual_2418477053.webp)
53
Limitations
The test strips are used for fresh capillary whole blood samples.
1. DO NOT use neonate blood sample.
2. Not to be used for diagnosis or screening of diabetes.
3. Alternative site testing with this system can be used only during steady-state blood glucose conditions.
4. Measurements from alternative site testing should never be used to calibrate a continuous glucose monitor
(CGM) or entered into insulin dose calculators for insulin dosing recommendations.
5. Extreme humidity may affect the results. A relative humidity greater than 90 % may cause incorrect results.
6. The system should be used at a temperature between 50°F and 104°F (10°C and 40°C). Outside this
range, the system may get incorrect results.
7. DO NOT reuse the test strips. The test strips are for single use only.
8. Hematocrit: The hematocrits between 20% and 60% will not significantly affect the results. Hematocrit
below 20% may cause higher results. Hematocrit above 60% may cause lower results. If you do not know
your hematocrit level, please consult with your healthcare professional.
9. Altitude up to 10,000 feet above sea level has no effect on readings.