Page 3
EasyMax LTC System
Specifications
Brand Name
EasyMax LTC Blood Glucose Meter
Range
20-600 mg/dL
Test Time
5 Seconds
Memory Sets
480 Test Results
Operating Condition
Temp.
50
0
F-104
0
F (10
0
C-40
0
C)
Relative Humidity
R.H.
<
90%
Blood Sample
0.6 µL
Fresh Blood from Fingertip
Hematocrit (Hct)
20~60%
Power
Alkaline Battery AAA (2 ct.)
Battery Life
Over 2,000 Tests
Display Dimension
1.4” x 1.7” (35 x 43 mm)
Device Dimension H x W x D
3.7” x 2.0” x 0.8” (94 x 50 x 19.5 mm)
Weight
1.68 oz (47.7 + Grams) w/o Batteries
Principles
Electrochemical Biosensor Technology
Software Via USB
GlucoManager™
Unusual Test Results
If the results are inconsistent with your patient’s symptoms, please follow these steps:
1. Run a control solution test (page 7).
2. Repeat a blood glucose test (page 9).
3. After running the quality control and retesting the patient, if there are still concerns about the
performance of the meter, inform the Nursing Supervisor and follow your facility’s procedure for
cross-checking results with the Laboratory or another glucose test method (not another meter).
4. For accuracy and precision data and for important information on limitations, see the instructions
that come with your test strips.
Note: 1. Extremely high humidity may affect the test results. A relative humidity above 90% may
cause inaccurate results.
2. Hematocrit below 20% may cause higher results. Hematocrit above 60% may cause
lower results.
=
-