EasyMax LTC System

Specifications

Brand Name

EasyMax LTC Blood Glucose Meter

Range

20-600 mg/dL

Test Time

5 Seconds

Memory Sets

480 Test Results

Operating Condition

Temp.

50

0

F-104

0

F (10

0

C-40

0

C)

Relative Humidity

R.H. 

<

 90%

Blood Sample

0.6 µL

Fresh Blood from Fingertip

Hematocrit (Hct)

20~60%

Power

Alkaline Battery AAA (2 ct.)

Battery Life

Over 2,000 Tests

Display Dimension

1.4” x 1.7” (35 x 43 mm)

Device Dimension H x W x D

3.7” x 2.0” x 0.8” (94 x 50 x 19.5 mm)

Weight

1.68 oz (47.7 + Grams) w/o Batteries

Principles

Electrochemical Biosensor Technology

Software Via USB

GlucoManager™

Unusual Test Results

If the results are inconsistent with your patient’s symptoms, please follow these steps:

1.  Run a control solution test (page 7).

2.  Repeat a blood glucose test (page 9).

3.  After running the quality control and retesting the patient, if there are still concerns about the 

  performance of the meter, inform the Nursing Supervisor and follow your facility’s procedure for 

  cross-checking results with the Laboratory or another glucose test method (not another meter).

4.  For accuracy and precision data and for important information on limitations, see the instructions 

  that come with your test strips.

Note:  1.  Extremely high humidity may affect the test results. A relative humidity above 90% may

      cause inaccurate results.

    2.  Hematocrit below 20% may cause higher results. Hematocrit above 60% may cause 

      lower results.

=

-

Содержание EasyMax LTC

Страница 1: ...ng Blood Glucose System This Manual is intended for facilities where point of care testing is conducted by a healthcare professional Please thoroughly read the EasyMax LTC Owner s Guide and all produc...

Страница 2: ...tured by EPS Bio Technology Corp No 8 R D Road III Hsinchu Science Park Hsinchu Taiwan 30077 Customer Service 1 866 994 3345 Toll Free Mon Fri 8 00 AM 6 00 PM EasyMax LTC Self Monitoring Blood Glucose...

Страница 3: ...Meter is designed and approved for testing fresh capillary whole blood samples from your fingertip palm and forearm The meter is for outside the body in vitro use It should not be used to diagnose di...

Страница 4: ...ic Acid 0 8 1 2 4 4 Dopamine NA 0 09 2 6 Gentisic Acid 1 8 1 8 6 Glipizide 0 2 0 2 8 Ibuprofen 0 5 4 2 50 50 L Dopa 0 5 8 0 2 Methyldopa 0 1 0 5 0 5 2 Sodium Salicylate 15 30 50 Tetracycline 0 4 1 5 1...

Страница 5: ...ts If the results are inconsistent with your patient s symptoms please follow these steps 1 Run a control solution test page 7 2 Repeat a blood glucose test page 9 3 After running the quality control...

Страница 6: ...rmicidal wipe Open unfold and use the first germicidal wipe to remove heavy soil Step 2 Disinfection Instruction After cleaning and before using the meter again unfold a new wipe and thoroughly wet al...

Страница 7: ...st result or error message E Insert a Test Strip Insert a test strip to start testing blood glucose F Medical Record No Shows the patient number G Apply Control Solution Control solution test marker H...

Страница 8: ...e Use Setting the current time and date on your meter is important if you use the meter memory 4 Repeat step 3 to set the date and time The flashing field is the one you are currently setting 5 Press...

Страница 9: ...mber you select if you flag the result as a control solution test The result will be displayed in the meter s memory as a control solution test Page 7 2 Press Left Button to select the mode of Control...

Страница 10: ...ol test with properly stored strips Was the test strip bottle capped tightly If it wasn t the strips could be humidified Replace the test strips Is the control solution expired or contaminated Replace...

Страница 11: ...firmly against the puncture site 7 Gently squeeze and or massage the fingertip until a round drop of blood forms on the fingertip Page 9 Note If the function of is on and you do not choose a medical...

Страница 12: ...f 500 F 1040 F 100 C 400 C This alerts users that an incorrect result may occur if the test continues Action Relocate the meter to a location with temperature between 500 F 1040 F 100 C 400 C The LF a...

Страница 13: ...EasyMax LTC Self Monitoring Blood Glucose System Training Information for Healthcare Professionals Page 11...

Страница 14: ...end of test strip Identifies the designated hand hold area 4 Demonstrates proper test strip use Only uses EasyMax LTC blood glucose meters with EasyMax 15 blood glucose test strips Runs a control solu...

Страница 15: ...Quality Control Performs a control solution test and can explain why it should be done Understands the dates listed on the control solution and how they apply to opened vs unopened vials Recognizes wh...

Страница 16: ...ill lower or higher than expected D All of the Above 4 When performing a blood glucose test where should you apply the blood A On top of the test strip B On the side of the test strip C The tip of the...

Страница 17: ...type of battery does the EasyMax LTC require A 3 Volt Lithium B Alkaline 1 5 Volt AAA C Alkaline 1 5 Volt AAAA D None of the above 12 Describe the proper way to dispose of a safety lancet 13 Describe...

Страница 18: ...Glucose Quality Control Results Log Facility Name Monitor Month Year Results Reviewed by Date Time Staff Performing Test Strip Lot Level 1 Control Level 2 Control Action Taken Lot Range Result Lot Ran...