Menu Flow Diagram
Menu can be accessed by using the MENU key by pressing the key continuously for 2 sec.
Calibration accesses through hardware lock protection. Please connect and short pin no. 2
and 3 in rear panel to enable calibration.
Two different Load calibrations available “AUTO” & “MANUAL”. Difference, in manual
calibration
User need to enter Known weight count (without zero count) as span
count.