Emissions test

RF emissions CISPR 11

Harmonic Emissions
IEC 61000-3-
Voltage Fluctuations/Flicker 
Emissions 
IEC 61000-3-

 
 
 

Immunity Test

Electrostatic discharge (ESD)
IEC 61000-4-
Electrical fast
transient/burst
IEC 61000-4-
Surge 
IEC 61000-4-
Voltage dips, short 
interruptions and voltage
variations on power supply 
input lines 
IEC 61000-4-
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-
Conducted RF
IEC 61000-4-
Radiated RF*
IEC 61000-4-

 
*Recommended separation distance
d = 1.17 

d = 0.35 

P 80 MHz to 800 MHz

d = 0.70 

P 800 MHz to 2.5 GHz

Where  (P)  is  the  maximum  output  power
transmitter  manufacturer  and  d  is  the  recommended  separation  distance  in  meters  (m).  Field 
strengths from fixed RF transmitters, as determined by an electromagnetic site survey, should be less 
than the compliance level in each frequency range.
Recommended distance between device and portable and mobile RF equipment.
 
 

EO

INTERFERENCE  MAY  OCCUR  IN  THE  VICINITY  OF  EQUIPMENT  MARKED  WITH  THE 

 

IS  DESIGNED  FOR  USE  IN  THE  ELECTROMAGNETIC  ENVIRONMENT  DESCRIBED 

BELOW. THOSE USING THE DEVICE SHOULD ENSURE THAT THE EO-BAT9

Emissions test 

Level of compliance 

RF emissions CISPR 11 

Class A 

Harmonic Emissions 

-2 

Class A 

Fluctuations/Flicker 

-3 

Complying 

Immunity Test

 

Level of compliance

 

Electrostatic discharge (ESD) 

-2 

6 kV contact 
8 kV air 

Electrical fast 
transient/burst 

-4 

1 kV 
 

-5 

1 kV differential mode 
2 kV common mode 

Voltage dips, short 
interruptions and voltage 
variations on power supply 

-11 

<5% Ut for 0.5 cycle 
40% Ut for 5 cycles 
70% Ut for 25 cycles 
<5% Ut for 5 sec 

Power frequency (50/60 Hz) 
magnetic field 

-8 

3 A/m 

Conducted RF 

-6 

3 Vrms 
150 kHz to 80 MHz 

Radiated RF* 

-3 

10 V/m 
80 MHz to 2.5 GHz 

*Recommended separation distance 

P 80 MHz to 800 MHz 
P 800 MHz to 2.5 GHz 

Where  (P)  is  the  maximum  output  power  rating  of  the  transmitter  in  watts  (W)  according  to  the 
transmitter  manufacturer  and  d  is  the  recommended  separation  distance  in  meters  (m).  Field 
strengths from fixed RF transmitters, as determined by an electromagnetic site survey, should be less 

he compliance level in each frequency range. 

Recommended distance between device and portable and mobile RF equipment.

EO-BAT9 - User Guide 

 

18 

INTERFERENCE  MAY  OCCUR  IN  THE  VICINITY  OF  EQUIPMENT  MARKED  WITH  THE 

THE  ELECTROMAGNETIC  ENVIRONMENT  DESCRIBED 

 IS USED IN SUCH AN 

rating  of  the  transmitter  in  watts  (W)  according  to  the 

transmitter  manufacturer  and  d  is  the  recommended  separation  distance  in  meters  (m).  Field 
strengths from fixed RF transmitters, as determined by an electromagnetic site survey, should be less 

Recommended distance between device and portable and mobile RF equipment. 

Содержание EO-BAT9

Страница 1: ...EO EO BAT9 User Guide 1 101 20 Rev F ...

Страница 2: ...tilator 12 4 2 Keyboard indicators 12 4 3 Indicators when the EO BAT9 is charging 12 4 4 Indicators when the EO BAT9 is powering the ventilator 13 4 4 1 Battery life indicator 13 4 4 2 Temperature fail indicator 14 4 5 Battery life indicator 14 5 EO BAT9 maintenance 15 5 1 EO BAT9 cleaning 15 5 2 Periodic safety checks 15 6 Product specifications 16 6 1 External charger electrical characteristics ...

Страница 3: ...view and adhere to all warnings and cautions contained in the User guide specific to the ventilator in use Warnings FIRE HAZARD WARNINGS NEVER EXPOSE BATTERIES TO DIRECT FLAMES WARNINGS BEFORE USING EQUIPMENT ALWAYS VERIFY THE PROPER FUNCTIONING OF THE BATTERY PACK BEFORE USING NEVER USE A BATTERY PACK OR ANY COMPONENTS OR ACCESSORIES THAT APPEAR TO BE DAMAGED IF ANY SIGNS OF DAMAGE ARE EVIDENT CO...

Страница 4: ...TH ALL APPLICABLE STANDARDS AND PROVIDE ELECTRICAL POWER CORRESPONDING TO THE AC VOL CHARACTERISTICS ON THE EXTERNAL CHARGER LABEL TO ENSURE VENTILATION CONTINUES UNINTERRUPTED ENSURE ALTERNATIVE POWER SOURCES ARE AVAILABLE AC POWER SOURCE SECOND BATTERY PACK OR AN AUXILIARY DC CAR ADAPTER BE PREPARED FOR THE POSSIBILITY OF POWER FAILURE BY HAVING AN ALTERNATIVE MEANS OF VENTILATION READY FOR USE ...

Страница 5: ...EO BAT9 User Guide 5 2 General information 2 1 Description of the EO BAT9 EO BAT9 EO BAT9 Front view Keyboard USB port ...

Страница 6: ...ting indicator EO EO BAT9 Rear view Keyboard view Discharge cable connector on the charger cable is the same as the one on the discharge cable Charge Discharge Battery life indicator button Alarm indicator Battery life indicator EO EO BAT9 User Guide 6 is the same as the one on the discharge cable Battery life indicator EO BAT9 connector ...

Страница 7: ...th European legal requirements Manufacturer Keep dry This side up Fragile Handle with care High and low temperature limitation High and low humidity limitation Recyclable EO Description Consult operating instructions indicator Battery life indicator International Protection Marking IEC standard 60529 Protection against ingress of water and foreign objects Should not be disposed of in household was...

Страница 8: ... 8 2 4 Indications for use The EO BAT9 provides an external battery which connects to the EO ventilators It is an external DC power source The battery pack is suitable for use in institutional home and portable settings ...

Страница 9: ...TING THE VENTILATOR TO THE BATTERY PACK IF THE BATTERY PACK HAS BEEN TRANSPORTED OR STORED AT A TEMPERATURE THAT 20 C FROM THE TEMPERATURE IN WHICH IT WILL BE OPERATING ALLOW THE BATTERY PACK TO STABILISE IN ITS OPERATING ENVIRONMENT FOR AT LEAST TWO HOURS PRIOR TO USE Charge Discharge Connector Line up the connector on the cable with the port on the battery pack onto the socket until an audible c...

Страница 10: ...harge Discharge Connector Line up the connector on the DC power cable with the port on the battery pack ector onto the socket until an audible click occurs The click indicates the connectors are locked together Connect the end of the AC power cable to the charger and the other end to a main power DO NOT ATTEMPT TO CONNECT THE EO BAT9 CHARGER WITH THE EO VENTILATOR NOT ATTEMPT TO CONNECT THE EO VEN...

Страница 11: ...ING THE PATIENT EO USB port or other device to be charged by the EO BAT9 when connected The discharge cable must be plugged in to allow USB connection charge THE TABLET CAN NOT BE CHARGED WHEN THE BATTERY PACK IS CHARGING DO NOT CONNECT MORE THAN ONE EXTERNAL DEVICE TO THE USB CONNECTOR OF THE DO NOT TOUCH THE USB SOCKET WHEN TOUCHING THE PATIENT EO BAT9 User Guide 11 when connected ED WHEN THE BA...

Страница 12: ...attery pack the ventilator runs on the EO BAT9 battery The AC DC power indicator LED on the ventilator s Ventilator s keyboard EO BAT9 powering the ventilator Keyboard indicators is connected to a ventilator or to the charger the keyboard is active When it is the keyboard is inactive Pressing the push button active the keyboard and display the Keyboard activated Keyboard not activated the EO BAT9 ...

Страница 13: ...een Charge complete Charger LED colour code When the external charger is not connected to the EO BAT9 the LED is green Please refer to to know the specifications and the characteristics of the external charger Indicators when the EO BAT9 is powering the ventilator Battery life indicator The battery life indicator displays the remaining battery power If all the four LEDs are lit the battery Only th...

Страница 14: ...e DC POWER LOSS alarm Temperature fail indicator When the battery pack is operating outside recommended environmental conditions the alarm indicator LED lights up and the battery life indicator LEDs remains lit Temperature fail The ventilator automatically switches to the ventilator s internal battery and an audible and visible alarm triggers on the ventilator When the battery temperature becomes ...

Страница 15: ...S FROM ENTERING THE DEVICE ALWAYS UNPLUG THE BATTERY PACK BEFORE CLEANING AND VERIFY THAT IT S DRY BEFORE PLUGGING IT BACK CLEAN ONLY EXTERIOR SURFACES OF THE BATTERY PACK Wipe the exterior surfaces of the EO BAT9 with a damp cloth using a mild cleaning solution as often y pack periodically whenever it is soiled or dirty and before storing Periodic safety checks The user should perform some safety...

Страница 16: ...tage output 22 VDC nominal Full load capacity 4 3 Ah Battery recharge time Full recharge 4 h Charging current 1 4 A 6 3 EO BAT9 battery run time The EO BAT9 run time is determined by Environmental conditions The conditions and age of the battery The ventilator settings The current circuit in place and unintentional leak The battery will operate for approximately nine hours when the ventilator is u...

Страница 17: ...ions could result in premature aging of the battery pack and a faster replacement Product compliance Class II Continuous IP2X IIb Guidance and manufacturer s declaration electromagnetic emissions and THE EO BAT9 SHOULD NOT BE USED IN CLOSE PROXIMITY TO OTHER EQUIPMENT OR STACKED ON TOP OF OTHER DEVICES IF THIS KIND OF USE IS UNAVOIDABLE THE EO BAT9 SHOULD BE CHECKED CAREFULLY AND OBSERVED TO ENSUR...

Страница 18: ...compliance RF emissions CISPR 11 Class A Harmonic Emissions 2 Class A Fluctuations Flicker 3 Complying Immunity Test Level of compliance Electrostatic discharge ESD 2 6 kV contact 8 kV air Electrical fast transient burst 4 1 kV 5 1 kV differential mode 2 kV common mode Voltage dips short interruptions and voltage variations on power supply 11 5 Ut for 0 5 cycle 40 Ut for 5 cycles 70 Ut for 25 cycl...

Страница 19: ... alkaline or other non acid electrolytes Safety requirements for portable sealed secondary cells and for batteries made from them for use in portable applications 6 8 Training and support Training and support are available on the EOVE website www eove fr or by calling our helpline at 33 0 5 59 21 86 84 6 9 Limited warranty The seller guarantees that the product delivered complies with the use for ...

Страница 20: ... Once defects or non conformity are duly recorded by the seller the purchaser may return the relevant product if the seller believes that it can be repaired in whole or in part Otherwise the seller shall substitute the non repairable dysfunctional equipment with equivalent new equipment In any case any return of products requested by the purchaser must be agreed in writing by the seller In particu...

Страница 21: ...EO BAT9 User Guide 21 2016 EOVE All right reserved Made in France 0459 LU LTXLGH HDOWKFDUH UHGLWYLHZ 5RDG 8QLW 0LVVLVVDXJD 2QWDULR 1 5 ZZZ DLUOLTXLGHKHDOWKFDUH FD ...

Отзывы: