Place the suction strainer (Foot Filter) in the
graduated cylinder.
The suction strainer will displace ~50 mL.
Note the volume in the graduated cylinder with
the suction strainer.
Using buttons, set the pump to
MANUAL MODE
.
Press button to enter the Mode Settings.
Press button, the screen should read
PULSES / MINUTE - PULSES NUM: 300
Press buttons to set
PULSES / MINUTE
to 100 (100 pulses / min).
Press the button, to save the new
PULSE /
MINUTE
settings.
Open
Test/Dump Port
to allow water flow.
GPM of water does not matter at this time, injection
will be done manually rather than proportionately.
Press the
START / STOP
button to activate the
pump. Run the pump for 1:00 Minute (use a timer).
3
4
5
6
250 mL
7
8
9
10
OPEN