Maintenance & Storage
Replace the batteries promptly when low voltage lamp is lighted
Clean surface of the oximeter before it is used in diagnosis for patients
Remove the batteries from the battery compartment when the oximeter is stored for a
long duration.
It is best to store the product in a place where ambient temperatures is –100C to
400C (140F -1040F) and humidity is 10%-80%
Store the oximeter in a dry environment anytime. A wet ambient might affect its
lifetime and even might damage the product.
1.
2.
3.
4.
5.
Calibrating the Pulse Oximeter
The functional tester cannot be used to assess the accuracy of the Oximeter.
The test methods used to establish the SpO2 accuracy is clinical testing.
The oximeter used to measure the arterial hemoglobin oxygen saturation levels and these
levels are to be compared to the levels determined from arterial blood sampling with a
CO-Oximeter.
Guidance and Manufacturer’s declaration - Electromagnetic Emission
The pulse Oximeter is intended for use in the electromagnetic environment specified
below. The user of the Oximeter should ensure that it is used as specified below.
Emission Test
Compliance
Electromagnetic Environment - Guidance
RF Emissions
CISPR 11
Group 1
The Pulse Oximeter uses RF energy for it’s internal
function only. Therefore, RF emissions from the
device is very low and are not likely to cause any
interference to other nearby equipment
RF Emissions
CISPR 11
Class B
The Pulse Oximeter is suitable for use in all
establishments, including domestic and those
directly connected to the public low-voltage power
supply networks that supply power to buildings
used for domestic purposes.
Possible Problems and Resolutions
Problem
Possible Reason
Solution
SpO2 or PR is not
appearing
Finger is not inserted
correctly
Oxy-hemoglobin is too
low to measure
Retry by inserting finger correctly
Try a few more times. If no display, please go to
a hospital or clinic immediately
SpO2 or PR is
unstable
Finger is not inserted
deep enough
Finger is shaking or user’s
body is moving
Retry by inserting finger deep enough
Try to stay still and measure again
The Oximeter is not
powering on
Battery must be
depleted or not present
Battery must be inserted
incorrectly
Please replace or insert new batteries
Insert batteries with polarities matching the
image shown in the battery compartment
The device may be
damaged
Please contact Customer Service
The indicator lamp
went off suddenly
Battery must be
depleted or not present
This is normal
Insert new Alkaline batteries
The device powers off
automatically when no
signal is detected for 8
seconds
“Error3” or “Error4”
displayed
Low Power
Receiving tube is
shielded or damaged or
connecter broken
Amp circuit malfunction
Insert new Alkaline batteries
Please contact Customer Support
“Error7”
displayed
Current control circuit
malfunction
Insert new Alkaline batteries
Please contact Customer Support
Emission tube damaged
Low Power
One year warranty from the date of purchase. Purchase invoice required. Warranty service shall not be
provided if the device malfunction is because:
1. The user opened the device and or dismantled and reinstalled the device
2. User inadvertently or otherwise dropped the device
3. lack of reasonable maintenance
4. User operated the device not in accordance with the guidelines provided in the user manual
5. Repair or maintenance done by an unauthorized personnel or establishment
Service provided outside the scope of warranty will be charged as per prevailing regulations and
stipulated rates. For warranty services and Customer Support please contact:
Eocene Systems,
2441 Bellevue Avenue,
Daytona Beach, FL 32114
Support: 321-567-0688 FAX: 321-567-0688
Warranty
SpO2
Symbol Definition
The equipment type is BF
Refer to User guide before application
Hemoglobin saturation; Heart rate BPM
Low Power indication