Duplication prohibited
MIR 9000
3–29
February 2000
For the analog inputs, it is necessary to specify the MUX channels used for acquisition (
MUX Channel
parameter). Channels 17 to 19 or DTP1 to DPT3 are reserved for this application. The connection is
then made direct to the interface 2 board.
The linearity parameters (
Ax2
,
Bx
and
C
) allow converting a voltage signal (0-2.5 volt) into a
measurement (0-200 °C for example).
CAUTION: The maximum voltage on these analog inputs is 2.5 volts.
The calculated values are determined from the analog inputs. Therefore, it is not necessary to specify
a MUX input or linearization.
For the pressures measured in mbar and the temperature in °C, the chimney flowrate is calculated in
KNm
3
/h (in thousandths of normal cubic metres per hour) and the speed in m/s (in metres per
second).
The units and display formats for each parameter should be configured in the “
Configuration /
Signal-Ranges-Units”
menu.
The lower part of the screen is common to each of the parameters.
The
K Pitot
coefficient corresponds to the coefficient of the Pitot tube itself in use (usual coefficient:
0.75).
The
Mass vol.
parameter corresponds to the volumetric mass of gas (usual coefficient: 1.293 g/l).
The
Chimney
parameter corresponds to the diameter of the chimney, in metres.
The
Backblow
parameter defines the frequency of back-blowings per hour (01h, 03h, 06h, 12h or
24h); the duration is defined by the
Time
parameter.
The measurements are blocked during back-blowing and a stabilisation period equal to the duration of
the back-blowing.
The back-blowing command is transferred to the EVC5 connection of the TIG2 board.
For a nil frequency (00 h), the measurements are blocked during each of the back-blowing cycles of
the MIR. This function enables back-blowing the Pitot tubes via a command of the SEC Probe.
Output
Unit:
MUX channel
Chimney flowrate
Fext – KNm
3
/h
Calculated value
-
Chimney temperature
Text - °C
Analog input
17
Chimney pressure
Pext – mbar
Analog input
18
Gas speed
Sext – m/s
Calculated value
-
Pitot pressure
Pdif – mbar
Analog input
19
Figure 3.3 – Configuration of the five DTP parameters
STOP
Содержание MIR 9000 - LCD
Страница 10: ...MIR 9000 Duplication prohibited 1 2 February 2000 9000 MIR Figure 1 1 Presentation of the MIR 9000 analyzer...
Страница 12: ...MIR 9000 Duplication prohibited 1 4 February 2000 Figure 1 2 Display and keyboard...
Страница 26: ...MIR 9000 Duplication prohibited 1 18 February 2000 Figure 1 12 Dimensions of the MIR tight boxes...
Страница 94: ...MIR 9000 Duplication prohibited 3 56 February 2000 Page left blank deliberately...
Страница 96: ...MIR 9000 Duplication prohibited 4 2 February 2000 Page left intentionally blank...
Страница 114: ...MIR 9000 Duplication prohibited 4 20 February 2000 Page left intentionally blank...
Страница 116: ...MIR 9000 Duplication prohibited 5 2 February 2000 Page left blank deliberately...
Страница 142: ...MIR 9000 Duplication prohibited 5 28 February 2000 Figure 5 19 Motherboard configuration...
Страница 144: ...MIR 9000 Duplication prohibited 5 30 February 2000 Figure 5 20 Test points and settings of the motherboard...
Страница 146: ...MIR 9000 Duplication prohibited 5 32 February 2000 Figure 5 21 Configuration and settings of the DAC8 board...
Страница 153: ......