background image

+1.800.343.5319 or +1.843.881.6560    envexp.com    

 

Environmental Express 

• 7

SimpleDist® Micro Heating Block:

 Operation & Instruction Manual   

 

Distillation Methods

Total Cyanide Distillation with the SimpleDist® Micro Heating Block

4.4   Sulfuric acid/magnesium chloride solution - Add 277 mL of reagent water to a 500 mL 

beaker. Add 80.5 g MgCl

2

•6H

2

O (4.3) and stir to completely dissolve. Slowly add 189 mL 

of sulfuric acid (4.2) while stirring. This should be done in a hood as toxic fumes may be 
released. Once the solution is cool you should have 500 mL of total volume.

4.5  Potassium cyanide 
4.6  Potassium hydroxide
4.7  Reagent water ASTM Type II or equivalent

Note: 

The toxicity for each of the reagents used in this procedure are not fully documented. 

Treat each chemical as a potential health hazard and limit exposure. Exercise good laboratory 
technique with an emphasis on safety.

5.0  Procedure (Note: Standards must be distilled in a like manner as the samples for this 
method)

5.1 

 Set the temperature of the Heating block to 120°C and allow the heat to stabilize. This 
should take approximately 30 minutes. If all reagents are ready to use this should be 
sufficient time to set up and prepare a full batch of 20 samples and all necessary QC.

5.2 

 Place one collector tube for each sample, standard, QC, or other aliquot needing to 
be distilled in a test tube rack with the “Measure” end up. Add 1.59 mL of 0.95 M 
NaOH solution (4.1) to each tube. 

Note:

 After final distillate dilution this will give a 

concentration of 0.25 M NaOH. The initial concentration may be adjusted to provide 
alternate final concentrations as needed by the analytical method of choice.

 Cap the 

tubes with one membrane and cap. Ensure that the membrane is not torn and shows  
a visible edge all the way around the tube.

5.3  

 Place one sample tube for each sample, standard, QC, or other aliquot needing to be 
distilled in a test tube rack and add 6.0 mL of sample/standard to each tube. The tubes 
may be labeled with permanent ink to aid in sample identification. Do not add any 
label to the tube as this will interfere with the fit in the block as well as heat transfer 
during the distillation process. If solid samples are being distilled add 0.5 to 1.0 g of 
the sample to the tube along with approximately 5 mL of reagent water. Samples high 
in organic content such as sludge should tend toward the lower sample mass while 
samples that are relatively free of organic material can use the higher end. The total 
amount of cyanide present in the tube should be less than 600 µg. 

5.4  

 Add 0.75 mL of the sulfuric acid/magnesium chloride solution (Step 4.4) to the first 
sample tube. Immediately place the ‘Distill’ end of a filled and capped collector tube 
(Step 5.2) over the open end of the sample tube. Press the two pieces together to form 
a rough fit. You will not be able to hand press a tight seal.

5.5  

 Place the distillation unit in the tube press. Hold the collection tube by the break joint 
to give support during the sealing procedure. Firmly press the two pieces together with 
one motion. The collection tube should be flush with the ring around the edge of the 
sample tube. If the pieces are at a visible angle the assembly will not fit all the way in 
the block. 

5.6  

Repeat Steps 5.4 and 5.5 for all prepared sample and collection tubes.

Содержание SimpleDist

Страница 1: ...SimpleDist Micro Heating Block Operation Instruction Manual 2021 Environmental Express All rights reserved July 2021 1065EE2_M Rev 2 1 800 343 5319 or 1 843 881 6560 envexp com...

Страница 2: ...lation Methods Total Cyanide Distillation 6 8 Ammonia Nitrogen Distillation 9 11 Total Phenol Distillation 12 14 Care Maintenance Care Maintenance 15 Circuitry Diagram 16 Troubleshooting Guide 17 19 R...

Страница 3: ...from use of or in conjunction with this product the maximum liability of Environmental Express whether by reason of breach of contract tort indemnification or otherwise but excluding liability of sel...

Страница 4: ...operate the SimpleDist Micro heating block in the vicinity of combustible material Users should be aware of potential dangers from heating certain types of compounds Such dangers may include explosion...

Страница 5: ...ment for the Heating of Materials CSA C22 2 61010 2 01 Note The SimpleDist Micro Heating Block is manufactured using the HotBlock technology and circuitry Information and Markings Environmental Expres...

Страница 6: ...0 4 Save original packaging material in a dry area for use if unit needs to be returned for service Refer to warranty policy on page 1 Installation Requirements Locate the SimpleDist Micro Heating Blo...

Страница 7: ...ture Settings Plug the SimpleDist Micro Heating Block into an approved receptacle Turn the heating block on by pressing the button on the back of the unit Wait until the display shows the current bloc...

Страница 8: ...i automated colorimetry manual colorimetry titrimetric or ion selective electrode 2 2 Reduced volume versions of this method that use the same reagents and molar ratios as in the original method are a...

Страница 9: ...adjusted to provide alternate final concentrations as needed by the analytical method of choice Cap the tubes with one membrane and cap Ensure that the membrane is not torn and shows a visible edge al...

Страница 10: ...quid into the disposal container Place the collection tube with the Measure end the capped end down in a test tube rack Repeat Steps 5 8 through 5 10 for all distilled tubes 5 11 Allow the tubes to co...

Страница 11: ...he distillate is then analyzed by one of the methods listed above 2 2 Reduced volume versions of this method that use the same reagents and molar ratios are acceptable provided they meet the quality c...

Страница 12: ...will depend upon the requirements of the analytical method Note After final distillate dilution this will give a concentration of 0 04 N H2SO4 or boric acid The initial concentration may be adjusted...

Страница 13: ...ple tube must be removed within 4 seconds of the unit being removed from the block Otherwise the distillate could suck back into the sample tube and cause the sample to be lost 5 10 Drop the sample tu...

Страница 14: ...irements 3 0 Interferences 3 1 Sulfur compounds can interfere with the process These compounds are removed by adjusting the pH of the sample to about 4 prior to distillation 3 2 Oxidizing agents can p...

Страница 15: ...of organic material can use the higher end 5 5 Immediately place the Distill end of a filled and capped collector tube Step 5 2 over the open end of the sample tube Press the two pieces together to fo...

Страница 16: ...roll the tube while tilting it back to level and then the Measure end down You may need to gently tap the tube to knock down any drops that cling to the edges 5 14 With the Measure end down snap the t...

Страница 17: ...g any cleaning or decontamination methods except those recommended check with Environmental Express to confirm the proposed method will not damage your heating block Avoid excessive spills as liquid a...

Страница 18: ...1 800 343 5319 or 1 843 881 6560 envexp com Environmental Express 16 SimpleDist Micro Heating Block Operation Instruction Manual Circuitry Diagram...

Страница 19: ...with 6 mL of water as a reference for the fill level Add one small boiling stone Hengar stones are preferred to the tube prior to adding the releasing agent and sealing the distillation unit Try not t...

Страница 20: ...illuminate and the above steps have passed the controller is faulty Call Environmental Express at 1 800 745 8218 for service There are two possible causes for your heating block to sound the alarm im...

Страница 21: ...ailure of the internal systems To reset default settings 1 Hold the up and down arrow buttons for six seconds until Ai Set appears 2 Hold the down arrow until glbl set is in the window 3 Press the gre...

Страница 22: ...Micro 240 VAC C6300 240 Power Module Plug Receptacle with Push Button Switch SC941 Power Cord Heavy Duty SC958 Environmental Express Controller SC945 110 220 V 25 A Solid State Relay SC952 Type K The...

Отзывы: