Section 6 -
Prior to Operation
The IAQ 15 Connect device comes with IAQ 15 App. The app can be downloaded from IOS and Android
Platform. Before use, the correct date, time, and user location should be verified and set.
1.
When the app is first used, user will need to go through the sign up and scan the device.
2.
To connect the app to the device, turn the IAQ 15 Connect pump on and once it is in main menu screen,
press the Bluetooth
®
button on pump. Bluetooth icon will appear next to battery level indicator and icon
will be gray.
3.
In the app, “IAQ 15” will appear on screen, select it and this will pair pump to your mobile device.
4.
Once paired, the Bluetooth icon on pump will turn from gray to blue.
5.
User can then use the app to:
a.
sync date and time
b.
sync location
Section 7 -
Calibration
It is recommended that IAQ 15 Connect is calibrated and verified at least once every day of use. A flow indicator
for Air-O-Cell cassettes is supplied with the pump for ease of calibration. You may also calibrate the pump using
an optional TSI
®
4046 Primary Calibrator with attachments or using an optional Via-Cell flow indicator (if
sampling with Via-Cell cassettes). These are the only approved calibrators that are specifically designed to work
with low backpressure impeller fan type pumps.
Warning:
DO NOT calibrate the pump using any other devices. Doing so will yield inaccurate readings due to the
backpressure created by the calibrators and may void warranty. Calibrators known by Environmental Express to
cause these inaccuracies include, but are not limited to, Bios Drycal, Bios Defender, Gilian Gilibrator, Gilian
Challenger, and standard MiniBuck calibrators.
Adjusting the Calibration
1.
Press the
On/Off
button to turn the pump on.
2.
Press
F2
button to select
Calibration
mode
from the main menu screen.
3.
Connect the flow indicator or the TSI 4046 Primary Calibrator to pump.
Air-O-Cell flow indicator connected to pump.
TSI 4046 Primary Calibrator connected to pump.