Environmental Express 35643-00 Скачать руководство пользователя страница 14

 

- 12 - 

Appendix – 1    Table of Parameter Setting

 

 
 

Appendix –2      Icons and Abbreviation

 

 

Appendix –3      Table of Self-diagnosis Symbol

 

Mode

 

Prompts

 

Parameter setting 

items

 

Abbreviation

 

Description

 

Factory 
Default

 

P4.0

 

DO

 

P4.1

 

Resolution

 

 

0.01-0.1(mg/L, ppm)

 

0.1-1 ( % )

 

0.01 mg/L

 

P4.2

 

Salinity calibration

 

 

0-45 ppt

 

0.01

 

P4.3

 

Air pressure 

adjustment

 

 

60-200 kPa

 

101.3

 

P4.4

 

DO unit

 

 

mg/L-ppm

 

mg/L

 

P4.5

 

Restore to factory 

default setting

 

 

No-Yes

 

No

 

P6.0

 

Basic 

Parameter

 

P6.1

 

Temperature unit

 

 

°C

 -

°F

 

/

 

P6.2

 

Automatic lock-up 

timing

 

/

 

Off-On

 

/

 

P6.3

 

Backlight timing

 

 

1-2-3-Off

 

/

 

P6.4

 

Automatic power-

off timing

 

 

10-20-30-Off

 

/

 

Mode

 

Prompts

 

Abbreviation

 

Stand for

 

Explanation

 

P4.0

 

DO

 

P4.1

 

 

Resolution

 

Resolution

 

P4.2

 

/

 

  Manual/Automatic

 

Manual or automatic 

salinity compensation

 

P4.3

 

 

Air pressure

 

Air pressure adjustment

 

P4.4

 

 

Unit

 

DO unit

 

P4.5

 

 

Factory default setting

 

Factory default setting

 

P6.0

 

Basic 

Parameter

 

P6.1

 

 

Unit

 

Temperature unit

 

P6.2

 

/

 

/

 

/

 

P6.3

 

 

Backlight setting

 

Backlight setting

 

P6.4

 

 

Automatic power-off

 

Automatic power-off 

timing setting

 

Icon

 

Self-diagnosis information

 

DO

 

 

  Press              when measuring value is not stable during calibration.

 

 

 

Solution error during salinity calibration; 

 

Salinity reading out of range of the meter recognition.

 

 

Содержание 35643-00

Страница 1: ...DO100 Portable DO Meter 35643 00 35643 05 Kit Instruction Manual ENVIRONMENTAL EXPRESS www environmentalexpress com ...

Страница 2: ...tment 5 5 Measurement of water samples 5 6 Notes of the measurement 5 7 Self diagnosis information 6 Parameter Setting 6 1 Main menu of parameter setting 6 2 Sub menu of DO parameter setting 6 3 Sub menu of basic parameter setting 7 What s in the Kit 8 Warranty Appendix 1 Table of Parameter Setting Appendix 2 Icons and Abbreviation Appendix 3 Table of Self diagnosis Symbol Appendix 4 Table of DO o...

Страница 3: ...atic power off and low voltage display The meter s digital filter improves measurement speed and accuracy There is a stable reading indication on the display Polarographic DO electrode comes with a special DO calibration sleeve taking only 3 5 minutes to polarize The electrode also comes with 3 combined membrane caps so that the membrane can be easily replaced The meter is dustproof and waterproof...

Страница 4: ...escription 3 1 LCD display 3 2 Keypad functions Table 1 Keypad operations and descriptions Measurement reading Parameter mode icon Units of measurement Temperature units Indicator icons Temperature value and indicator icons ATC automatic temperature compensation Automatic reading hold icon Stability icon of readings Low battery icon Diagram 2 Diagram 1 Short press 2 seconds Long press 2 seconds ...

Страница 5: ...isplay mode of measurement Refer to Diagram 3 and Diagram 4 for the display mode in the units of mg L and ppm Diagram 5 shows the display mode in measuring unit of referring to air saturation Diagram 6 shows the display mode When salinity value is 3 ppt salinity displays at the bottom right of LCD means that the automatic compensated salinity is 15 ppt Auto 3 3 2 Display mode of stable reading Whe...

Страница 6: ...the backlight will be turned on all the time 3 4 3 Battery The meter adopts 3xAA batteries Please use LR6 alkaline battery to ensure the battery quality Battery life 500 hours without backlight When LCD displays icon please replace with new batteries see Diagram 8 4 Polarographic DO Dissolved Oxygen Electrode 4 1 Structure of the DO electrode The meter connects to DO500 polarographic DO electrode ...

Страница 7: ...he cathode gold chip with a bit of force 4 4 2 Inject the electrolyte into a new membrane cap slowly make sure no air bubbles is generated Flick the membrane cap with fingers to eliminate the bubbles Then put the membrane cap on the table and slowly screw in the electrode clockwise vertically until the membrane cap is firmly screwed on At this time the excess electrolyte will be squeezed out wipe ...

Страница 8: ...n saturation is 100 and the concentration is 8 25mg L 77 F 25 C which means that the measured value in the air is 10 greater than it in water Therefore the meter is only suitable to calibrate in air with calibration sleeve and then measure the dissolved oxygen in water 5 4 Other calibration and adjustment 5 4 1 Zero oxygen calibration Zero oxygen calibration is generally only needed when replacing...

Страница 9: ...right of LCD Immerse the electrode into 12 88mS cm conductivity standard solution Note that the solution should exceed the height of the salinity electrode Stir gently and wait until the salinity value becomes stable and the stable icon displays on LCD press to calibrate see Diagram 10 b press twice to return to measurement mode see Diagram 10 c 5 4 3 Adjust the air pressure value The meter has th...

Страница 10: ... electrode shell and it contacts with the water directly Due to the different sensitivity between the thermistor and the electrolyte liquid it usually takes 3 to 5 minutes before the temperature of the electrolyte reaches the temperature of the temperature sensor For this the reading period must be 3min otherwise a large error will occur Especially when there s a big difference between the electro...

Страница 11: ...4 2 P4 5 P4 1 Resolution 0 01 0 1 1 Press 0 01 flashes press to select resolution press confirm The meter automatically displays 0 1 or 1 corresponding to 2 Press to enter P4 2 or press to return P4 2 Salinity calibration Refer to Section 5 4 2 for the salinity calibration steps Submenu of DO parameter setting Sub menu of basic parameter setting P4 1 Resolution P4 2 Salinity calibration P4 3 Air p...

Страница 12: ...press or to switch among P6 1 P6 2 P6 3 P6 4 P6 1 Temperature unit C F 1 Press C flashes press to select C F press to confirm 2 Press to enter P6 2 or press to return P6 2 Automatic lock up Off On 1 Press OFF flashes press to select OFF On press to confirm Off Unlock up On lock up Stable reading 10 seconds and the reading will be automatically locked up 2 Press to enter P6 3 or press to return P6 ...

Страница 13: ...oduct attributable to responsibility of OAKTON INSTRUMENTS for a period of THREE YEARS SIX MONTHS for the probe from the delivery This limited warranty does not cover any damages due to External causes such as accidents abuse improper use Normal wear and tear Failure to follow the product instructions or to perform any preventive maintenance modifications combination or use with any products mater...

Страница 14: ...ng Off On P6 3 Backlight timing 1 2 3 Off P6 4 Automatic power off timing 10 20 30 Off Mode Prompts Abbreviation Stand for Explanation P4 0 DO P4 1 Resolution Resolution P4 2 Manual Automatic Manual or automatic salinity compensation P4 3 Air pressure Air pressure adjustment P4 4 Unit DO unit P4 5 Factory default setting Factory default setting P6 0 Basic Parameter P6 1 Unit Temperature unit P6 2 ...

Страница 15: ...3 2 34 0 7 07 51 8 11 0 11 01 73 4 23 0 8 57 95 0 35 0 6 95 Appendix 5 Table of DO of Saturated Water at Different Altitudes Altitude Barometric pressure DO Altitude Barometric pressure DO Feet Meter kPa mmHg mg L Feet Meter kPa mmHg mg L 0 0 101 3 760 8 25 7500 2287 77 1 579 6 28 500 152 99 34 746 8 09 8000 2439 75 63 568 6 16 1000 305 97 6 733 7 95 8500 2591 74 44 559 6 06 1500 457 95 87 720 7 8...

Страница 16: ...6 762 101 60 10 11 8 27 6 97 763 101 73 10 12 8 28 6 98 764 101 86 10 14 8 30 6 99 765 102 00 10 15 8 31 7 00 766 102 13 10 16 8 32 7 01 767 102 26 10 18 8 33 7 02 768 102 40 10 19 8 34 7 02 769 102 53 10 21 8 35 7 03 770 102 66 10 22 8 36 7 04 771 102 80 10 23 8 37 7 05 772 102 93 10 25 8 39 7 06 773 103 06 10 26 8 40 7 07 774 103 19 10 28 8 41 7 08 775 103 33 10 29 8 42 7 09 Conversion between m...

Страница 17: ...ronmental Express 2345A Charleston Regional Parkway Charleston South Carolina 29492 Toll Free 800 343 5319 Technical Support 800 745 8218 Email tech environmentalexpress com Website www environmentalexpress com ...