background image

             

 

                   

ION 100 Portable Ion Meter 

 

Instruction Manual

 

35613-65, -68

 

 

 

 

Environmental Express

 

www.environmentalexpress.com

 

 

Содержание 35613-65

Страница 1: ...ION 100 Portable Ion Meter Instruction Manual 35613 65 68 Environmental Express www environmentalexpress com...

Страница 2: ...nformation regarding pH calibration and measurement 9 4 6 Customized calibration take 1 60pH 6 50pH as example 10 4 7 Notes regarding customized calibration 10 4 8 Self diagnosis information 10 4 9 Ma...

Страница 3: ...der to use and maintain the instrument properly please read the manual thoroughly before use To improve instrument performance constantly we reserve the right to change the manual and accessories with...

Страница 4: ...86 9 18 and 12 45 pH NIST mV Range 1000mV Resolution 200 200mV 0 1mV Other range 1mV Accuracy 0 2 F S Ion Range pX 0 00 10 00 Ion concentration 0 1999 Switchable unit pX mol L ppm mg L Resolution 3 4...

Страница 5: ...g ATC automatic temperature compensation MTC manual temperature compensation Ion symbol Calibration guide icon Units of measurement Automatic reading hold icon Temperature units Stability icon of read...

Страница 6: ...rt press In measurement mode press to turn on off the backlight In calibration mode press to calibrate In the parameter setup mode press to enter and open the sub menu confirm parameter When measureme...

Страница 7: ...to change the temperature value long press for quick changing of the temperature value 3 5 2 Automatic Power off The meter is with automatic power off function Select from parameter P6 4 to select aut...

Страница 8: ...rectly goes to automatic temperature compensation mode When the meter is connected to the pH electrode only it will enter manual temperature compensation mode Press or to input the temperature into th...

Страница 9: ...H7 00 buffer to conduct the 1st point of calibration Dip the pH electrode and the temperature probe into pH7 00 buffer solution stir gently and let it stand still and wait for the reading to be stable...

Страница 10: ...adopt 4 5 4 During the process of calibration and measurement the meter has self diagnosis functions indicating information Er1 Er4 please refer to Table 4 for relevant solutions 4 5 5 The calibratio...

Страница 11: ...into 1 60pH calibration solution stir gently and let it stand still and wait for the reading to become stable When the smiley icon displays on LCD press to adjust the value to 1 60 then press to confi...

Страница 12: ...ce The soaking solution contained in the supplied protective bottle is used to maintain hydration of the glass bulb and liquid junction Loose the cap take out the electrode and rinse in distilled or d...

Страница 13: ...ss bulb Electrodes that have been used for a long period of time will become aged Submerge the electrode in 0 1 mol L hydrochloric acid for 24 hours then rinse the electrode in distilled water then su...

Страница 14: ...quality or measuring value use ORP standard solution to test mV and see whether ORP electrode or meter works properly 5 3 2 Clean and activate ORP electrode After the electrode has been used over a lo...

Страница 15: ...ing should be 1 00 x 10 3 mol L When we use the unit of pX and ppm the reading we get is the value see Diagram 7 and 9 6 1 3 Choose the ion type The meter is built in with 16 ions frequently being use...

Страница 16: ...m right corner of the LCD When the reading is stable and the stable ion displays on LCD press or to adjust the value to be 1 00 E 5 1 00 10 5 press to finish the 1st point calibration the bottom left...

Страница 17: ...o be rinsed and the solution contamination can be avoided 6 3 3 Calibration error reminder Ion electrode calibration may fail often The reason could be calibration solution error the quality problem o...

Страница 18: ...4 Factory default setting For factory default setting please refer to parameter setting P5 3 All calibration data is deleted and the meter restores to the theory value Some functions restore to the or...

Страница 19: ...n menu and P6 0 for basic parameter menu Diagram 13 Main menu of parameter setting Sub menu of pH parameter setting Sub menu of ion parameter setting Sub menu of basic parameter setting P1 1 pH buffer...

Страница 20: ...ss to confirm USA USAseries nIS NIST series CUS Customized calibration 2 Press to enter P1 2 or press toreturn P1 2 Resolution 0 01 0 1 1 Press and 0 01 flashes press to select between 0 01 0 1 Press...

Страница 21: ...er the ion selection interface see Diagram b 2 Press or to move the cursor press to confirm 3 Press to enter P5 2 or press to return P5 2 Select between ppm and mg L 1 Press ppm flashes press to selec...

Страница 22: ...press to confirm Off Unlock up On lock up Stable reading 10 seconds and the reading will be automatically locked up 2 Press to enter P6 3 or press to return P6 3 Select backlight timing 1 2 3 Off 1 P...

Страница 23: ...ttributable to responsibility of Environmental Express for a period of THREE YEARS SIX MONTHS for the probe from the delivery This limited warranty does not cover any damages due to Transportation sto...

Страница 24: ...e to factory default setting No Yes No P6 0 Basic Parameter P6 1 Temperature unit P6 2 Automatic lock up timing Off On P6 3 Backlight timing 1 2 3 Off P6 4 Automatic power off timing 10 20 30 Off Appe...

Страница 25: ...ort 800 745 8218 Email tech environmentalexpress com Website www environmentalexpress com Icon Self diagnosis Information pH Ion Press when measuring value is not stable during calibration Wrong pH bu...

Отзывы: