PORTABLE SYNGAS ANALYZER IR-GAS-600P
10
3 Operation
3.1 Preparation before operation
1) Sample gas should connect to gas pretreatment system before using. The sample gas inletting into the gas
analyzer should be clean, dry and without tar.
2) Gas outlet pipe should be leaded to outdoor; Gas outlet pipe should not no blocked.
3) Sample gas pressure should be a little bigger than atmosphere pressure
4) Gas flow into the gas analyzer should control at 0.7L-1.2L/min, recommended to at 1.0L/m
3.2 Power on
1) Power on: press power button for more than five seconds until the display is activated. Here is showing
startup interface.
2) After startup, it shows MEASURE interface. (The picture is only for reference).
3.3 Measurement screen